China Anesthesia Residency Programs, Find details about China Medical Equipment, Anesthesia System from Anesthesia Residency Programs
Application:
The Anesthesia machine makes a good performance in Intensive Care Units (ICU), Operation room, Anesthesiology Department and other departments. From high to low acuity, simple to complex cases, pediatric to geriatric patients, Superstar Medical Anesthesia Systems offer you the choices you need in ventilation, monitoring and technique. what's more, our expertise in anesthesia delivery and ventilation is based on rich 23-year history of developing anesthesia solutions that meet your needs. Good performance based on High Security, High Accuracy, High Stability and accurate monitoring.
Features
1.Simplicity: four static casters with self-locking function.
2.Precision in an anesthesia ventilator with multiple ventilation modes: IPPV, A/C, SIMV and MANU.
3.10.4" TFT LCD screen displays the Ventilation parameters, Alarm information and Oscillographs.
4.Vaporizer with temperature, flow compensation and self-lock function, keep safety anytime.
5.Pressure-time, low-time and high precision ETCO2, O2 concentration show in real time.
6.ETCO2 and Anesthesia Gas Scavenging System (AGSS) are optional.
7.Built-in backup battery provide the emergency power supply to the unit.
8.Low O2 pressure alarm and N2O cut-off protection.
9.Over 5,000 units are installed in more than 200 countries.
10.Designed and manufactured by Superstar Medical with over 23 years of experience in ICU field.
Specifications
Ventilation modes: | IPPV; A/C; SIMV |
Graphical display: | |
P-T(pressure - time) | |
F-T(flow - time) | |
Alarm and protection | |
The AC power failure alarm | Power failure or no connection |
Internal battery backup low voltage alarm | <11.3±0.3V |
No tidal volume | ≤5Ml within 6s |
High oxygen concentration alarm Low oxygen concentration alarm | 19%-100% 18%-99% |
High Airway pressure alarm Low Airway pressure alarm High Minute Volume alarm Low Minute Volume alarm Continuous Pressure alarm | 20cmH2O-100cmH2O 0cmH2O-20cmH2O Adult(5L/min-20L/min) Paed(1L/min-15L/min 0-10L/min) (PEEP+1.5kPa) over 16s |
Suffocation warning | 5s-60s no spontaneous ventilation |
The maximum limited pressure | <12.5 kPa |
Fan error Oxygen deficit | Show on screen Show on screen |
Working conditions | |
Gas source | O2,N2O,AIR |
Pressure | 280kPa-600kPa |
Voltage | 100-240V |
Power frequency | 50/60Hz |
Input power | 80VA |
Vaporizer |
Anesthesia gas | Adjustable scope % (volume percentage) |
Halothane | 0 ~ 5 |
Enflurane | 0 ~ 5 |
Isoflurane | 0 ~ 5 |
Sevoflurane | 0 ~ 8 |