China Microbiology Laboratory Equipment Water Treatment System for Ultra Pure Water, Find details about China Microbiology Laboratory Ultra Pure Water, Microbiology Laboratory Water Equipment from Microbiology Laboratory Equipment Water Treatment System for Ultra Pure Water
Technical Parameter
Model | OK-MP10 | OK-MP20 | OK-MP40 | OK-MP60 |
Output | 10L/h | 20L/h | 40L/h | 60L/h |
Feed water requirements | Tap water,temperature 10~35ºC,pressure:0.15-0.35Mpa; TDS:≤200mg/L | |||
RO water quality |
| |||
Conductivity(25ºC) | 3-6μs/cm | |||
Iron rejection rate | 99%(new RO membrane) | |||
Organic rejection rate | >99%,when MW>200Balton | |||
Paeticles and bacteria rejection rate | >99% | |||
Ultrapure water quality |
| |||
Resistivity(25ºC) | 18.2MOhm-cm | |||
Bateria | <1CFU/ml | |||
Particle | >0.22um/ml<1um/ml | |||
AC requirement | 110/220V,50/60Hz | |||
Dimension(L*W*H-mm) | 410*300*500mm | |||
Weight(kgs) | 20kgs,30kgs,50kgs,70kgs | |||
Standard configuration | Main body,pressure tank |
The water purifying technology process is divided into source water pretreatment system,reverse osmosis system and ultra-pure system.
1. Source water pretreatment system: PP fiber filter + AC filter + Softener
2. Reverse osmosis system: It has high removal rate more than 99% for high valent ions, colloids, bacteria and organic matter (including endotoxin) with molecular weight greater than 300Dalton.
3. Ultra pure system: Ion exchange resin using H + exchange cation out the water in the filter ,using OH- exchange anions out the water , the replaced H + and OH- combined to produce H2O, ultra-purified column filling the United States Dow UPW-450 Ultra-pure water for nuclear grade resin, the water resistance of up to 18.2MΩ.cm.