Coating Thickness Tester Cm260

China Coating Thickness Tester Cm260, Find details about China Coating Thickness Meter Cm260, Coating Thickness Gaguge Dr260 from Coating Thickness Tester Cm260

Model NO.
CM260
Interface
RS-232c
Operating Condition
Temp. 0~50c
Range
0-1250/3000μm
Trademark
Beifon
Transport Package
Carton
Specification
ISO
Origin
China
HS Code
9026100000
Model NO.
CM260
Interface
RS-232c
Operating Condition
Temp. 0~50c
Range
0-1250/3000μm
Trademark
Beifon
Transport Package
Carton
Specification
ISO
Origin
China
HS Code
9026100000
Coating Thickness Meter CM8856/CM260/DR260

FEATURES
* It meets the standards of both ISO2178
And ISO 2361 as well as DIN, ASTM
And BS. It can be used both in the
Laboratory and in harsh field conditions.

* The F probes measure the thickness of
Non-magnetic materials (e. G. Paint,
Plastic, porcelain enamel, copper, zinc,
Aluminum, chrome etc. ) on magnetic
Materials (e. G. Iron, nickel etc. ). Often
Used to measure the thickness of
Galvanizing layer, lacquer layer,
Porcelain enamel layer, phosphide
Layer, copper tile, aluminum tile, some
Alloy tile, paper etc.

*The N probes measure the thickness of
Non-magnetic coating. Used on
Anodizing, varnish, paint, enamel,
Plastic coatings, powder, etc. Applied
To aluminum, brass, non-magnetic
Stainless steel, etc.



* Digital backlit display gives exact
Reading with no guessing or errors.

* Can store 99 groups of measurements.
Product ModelCoating Thickness Meter
Technical Specification
Operation languageEnglish
StandardsDIN, ISO, ASTM, BS
CalibrationZero and foil calibration
StatisticsNumber of measurements, mean, standard deviation,
maximum and minimum of 3000 readings
Data memory495 readings
LimitsAdjustable with alarm
InterfaceRS-232
Working temperature0-40
power supplyNicd rechargeable batteries 1.25V
Size270mm× 86mm× 47mm
WeightApprox. 530g
Standard deliveryMain unit
Calibration foil set
Substrate
Charger
Probe
Instruction manual
TIME certificate
Warranty card
Optional accessories6 optional probes
PC software Dataview
Calibration foils in different thickness
Connecting cable

 
    Features
*  Two measuring methods: magnetic induction (F) and eddy current (N)
*  Six types of probes are available for various applications( F400F1F1/90° F10N1CN02)
*   Two measurement modes: continuous / single
*   Five statistical ways: average, maximum, minimum, testing frequency, standard deviation
*   Two modes: direct mode, group mode
*   Have a delete function, can delete the single date or all the data
* Has a print function
* Two switch off modes: manual and auto