China Contec Cms8000 Hospital Patient Monitoring Equipment ICU Multiparameter Patient Monitor, Find details about China Patient Monitor, Multiparameter Patient Monitor from Contec Cms8000 Hospital Patient Monitoring Equipment ICU Multiparameter Patient Monitor
ECG | Multi-parameter Portable Cardiac Patient Monitor FDA |
Lead selection: I, II, III, aVR, aVL, aVF, V | |
Waveform: 5-lead, dual-channel | |
3-lead, single-channel | |
Gain: 2.5mm/mV, 5.0mm/mV, 10mm/mV, 20mm/mV, 40mm/mV | |
Scan speed: 12.5mm/s, 25 mm/s, 50 mm/s | |
HR: | Measurement and alarm range: 15~350 bpm |
Accuracy: ±1 % or ±1 bpm, whichever is greater | |
Alarm accuracy: ± 2 bpm | |
Resolution: 1 bpm | |
ST-segment monitoring: | |
Measurement and alarm range: -2.0 mV ~ +2.0 mV | |
Accuracy: -0.8 mV~+0.8 mV ±0.04 mV or ±10%, whichever is greater | |
Other range: unspecified | |
Arrhythmia analysis: ASYSTOLE, VFIB/VTAC, COUPLET, BIGEMINY, TRIGEMINY, R ON T, VT>2, | |
PVC, TACHY, BRADY, MISSED BEATS, PNP, PNC | |
Pacemaker: yes | |
RESP | Method: R-F(RA-LL) Impedance |
Respiration rate:Measurement and alarm range: 0~150 rpm | |
Resolution: 1 rpm | |
Measurement accuracy: ±2 rpm | |
Alarm accuracy: ±3 rpm | |
Apnea alarm: 10~40s | |
Scan speed: 6.25 mm/s, 12.5 mm/s, 25 mm/s | |
NIBP | Method: Oscillometry |
Mode: Manual/AUTO/STAT | |
Measurement interval in AUTO mode: 1/2/3/4/5/10/15/30/60/90/120/240/480/960 minutes | |
Measurement period in STAT mode: 5 minutes | |
Measurement and alarm range: 10 ~ 270 mmHg | |
Resolution: 1 mmHg | |
Cuff pressure accuracy: ±3 mmHg | |
Measurement accuracy: Maximal mean deviation: ±5 mmHg Maximal standard deviation: 8 mmHg | |
Over-pressure protection: Adult mode: 297±3 mmHg Pediatric mode: 240±3 mmHg Neonatal mode: 147±3 mmHg | |
Spo2 | Measurement and alarm range: 0 ~ 100% |
Resolution: 1% | |
Measurement accuracy: 70%~100%: ±2%; | |
0%~69%: unspecified | |
PR | Measurement and alarm range: 30 ~ 250 bpm |
Measurement accuracy: ±2 bpm or ±2%, whichever is greater | |
TEMP | Channel: dual-channel |
Measurement and alarm range: 0 ~ 50ºC | |
Resolution: 0.1 ºC | |
Accuracy: ±0.1 ºC | |
EtCO2 | Method: Sidestream or Mainstream Measuring Range: 0~150mmHg |
Resolution:0~69 mmHg, 0.1 mmHg 70~150 mmHg, 0.25 mmHg | |
Accuracy:0~40 mm Hg ±2 mm Hg 41~70 mm Hg ±5% 71~100 mm Hg ±8% 101~150 mm Hg ±10% | |
AwRR Range: 2~150 rpm | |
AwRR Accuracy: ±1BPM | |
Apnea Alarm: Yes | |
IBP | Channel: dual-channel |
Label: ART, PA, CVP, RAP, LAP, ICP, P1, P2 | |
Measuring and Alarm Range: -10~300 mmHg | |
Resolution 1 mmHg | |
Accuracy: ±2% or 1mmHg, whichever is greater | |
Power supply: 100-240V~, 50/60Hz | |
Safety classification: Class I, type CF defibrillation-proof applied part |