China 500va Acw/Dcw/IR Testing An9636hc Electrical Safety Testing Machine, Find details about China Electrical Safety Test, Electrical Safety Analyzer from 500va Acw/Dcw/IR Testing An9636hc Electrical Safety Testing Machine
The new-generation Ainuo comprehensive safety analyzer is mainly used for electrical safety testing of quality inspection institutions, laboratories and 3C and peripheral equipment. It has flexible channel configuration options and special design for power supply equipment, meeting requirements for multi-station quick test
Function | Technical indicator | ||
AC withstand voltage test | Nominal output capacity | 200VA (5000V/40mA), short circuit current is greater than 200mA; optional 500VA model. | |
Output voltage setting | Range/Accuracy | Range: 100~5000V, resolution: 1V, Accuracy: ±(1% × setting +5V) | |
Output frequency setting | Range/Accuracy | 50Hz/60Hz, Accuracy: ±0.1Hz | |
Alarm current setting | Upper limit/Accuracy | Range: 0.00~100.00mA, resolution: 0.01mA, Accuracy: ±(1% × reading + 5 counts) | |
Lower limit/Accuracy | Range: (0.000~9.999)mA, resolution: 0.001mA, Accuracy: ±(1% × reading + 5 counts) | ||
Time setting | Range of test time | Range: 0, (0.5~999.9)s, 0 is infinite, resolution: 0.1s, Accuracy: ±(0.1% × setting + 2 counts) | |
Ramp-up time | Range: 0, (0.1~999.9)s,0 is off, resolution: 0.1s, Accuracy: ±(0.1% × setting + 2 counts) | ||
Ramp-down time | Range: 0, (0.1~999.9)s,0 is off, resolution: 0.1s, Accuracy: ±(0.1% × setting + 2 counts) | ||
Initial voltage setting | (0%-50%) × setting of output voltage | ||
Current compensation setting | 0.000~10.00mA, total current + compensation current <100.00mA, automatic/manual | ||
Arc detection | 1~9 (9 corresponds to most sensitive), 0 means that arc function is Off | ||
DC withstand voltage test | Nominal output capacity | 120VA( 6000VDC/20mA ) | |
Output voltage setting | Range/Accuracy | Range: (100~6000) VDC, resolution: 1V, Accuracy: ±(1% × setting + 5V) | |
Alarm current setting | Upper limit/Accuracy | Range: 0.0~20000μA, resolution: 0.1μA/1uA, Accuracy: ±(1% × reading + 5 counts) | |
Lower limit/Accuracy | Range: (0.0-999.9)μA, resolution: 0.1μA, Accuracy: ±(1% × reading + 5 counts) | ||
Time setting | Range of test time | Range: 0, (0.5~999.9)s, 0 is infinite, resolution: 0.1s, Accuracy: ±(0.1% × setting + 2 counts) | |
Ramp-up time | Range: 0, 0.4~999.9s, 0 is off, resolution: 0.1s, Accuracy: ±(0.1% × setting + 2 counts) | ||
Ramp-down time | Range: 0, (1.0~999.9)s, 0 is off, resolution: 0.1s, Accuracy: ±(0.1% × setting + 2 counts) | ||
Initial voltage setting | (0%-50%) × setting of output voltage | ||
Lower limit current for charging | (0~350.0)uA, automatic/manual | ||
Arc detection | 1-9 (9 is the most sensitive), 0 means arc function off | ||
Insulation resistance test | Nominal output | 2500VDC/50GΩ | |
Output voltage setting | Range/Accuracy | Range: (100~2500) VDC, resolution: 1V, Accuracy: ±(1% × setting + 5V) | |
Alarm resistance setting | Range of upper/lower limit | Range: 0.10MΩ~50000MΩ, the upper limit includes the setting for no upper limit | |
Accuracy | 100~499V:0.10MΩ~999.99MΩ, ±(2% × reading + 2 counts) 1000~2000MΩ,±(5% × reading + 2 counts) 500~2500V: 0.10~999.99MΩ, ±(2% × reading + 2 counts) 1000~9999MΩ,±(5% × reading + 2 counts) 10000~50000MΩ,±(15% × reading + 2 counts) | ||
Time setting | Range of delay time | Range: 0, (0.5~999.9)s, 0 is infinite, resolution: 0.1s, Accuracy: ±(0.1% × setting + 2 counts) | |
Ramp-up time | Range: 0, (0.1~999.9)s, 0 is off, resolution: 0.1s, Accuracy: ±(0.1% × setting + 2 counts) | ||
Ramp-down time | Range: 0, (1.0~999.9)s, 0 is off, resolution: 0.1s, Accuracy: ±(0.1% × setting + 2 counts) | ||
Lower limit current for charging | (0~3.500)μA, automatic/manual | ||
Ground bond resistance test | Rated output | Test current up to 32A, open circuit voltage is lower than 12V; | |
Output current setting | Range/Accuracy | Range: (2.0A-32.0A) AC, resolution: 0.1A, Accuracy: ±(1% × setting + 2 counts) | |
Output voltage setting | Range/Accuracy | Range: 3.0~10.0 VAC, resolution: 0.1V, Accuracy: ±(1% × setting + 2 counts), open circuit | |
Output frequency setting | Range/Accuracy | 50Hz/60Hz, Accuracy: ±0.1Hz | |
Alarm limit setting | Range of resistance upper/lower limit | 3 A≤ output current ≤10A: 10~600 mΩ 11A≤ output current ≤25A: 10~300 mΩ 26A≤ output current ≤32A: 10~200 mΩ | |
Accuracy | <100mΩ, ±(1% × reading + 1mΩ); ≥100mΩ, ±(1% × reading + 2 counts) | ||
Compensation resistance | Range | 0~100mΩ, automatic measurement, compensation on/off | |
Test time setting | Range/Accuracy | Range: 0, (0.5-999.9)s, 0 is infinite, resolution: 0.1s, Accuracy: ±(0.1% × setting + 2 counts) | |
Dimension | 426*132*470(W×H×D,mm) |