TMAX-Verticalfurnacevertical split tube furnaces feature the famous  (Sweden) spiral wire coils embedded in Mitsubishi(Japan) alumina fiber insulation. A double walled steel internal housing helps minimize heat loss to exterior surface.
 
Furnace operation is controlled by Shimaden (Japan) 40-segment digital controller with built-in RS485 digital communications port and USB adaptor, allowing the user to connect to a PC for remote control and monitoring of the furnace. You can also save or export test results. All our furnaces are CE compliant.

Related products:

Quenching Furnace

Lab Single-Zone 1200c Vertical Tube Furnace

Features
1.      Compact, portable and lightweight design.
2.       Split furnace chamber for fast and easy tube access.
3.      Standard package includes set of stainless steel vacuum sealing flanges with valves and pressure meter.
4.      Microprocessor based self-tuning PID control provides optimum thermal process with minimal overshoot.
5.      Multiple atmosphere processes in a single cycle are possible (ex: binder burn out in air and parts sintering under rough vacuum or inert gas environment.)
6.       Built-in ammeter and dual voltmeters for easy monitoring and troubleshooting.
7.       Built-in computer interface.
8.       Long life type K thermocouple.
 
Safety
1.       Overheat protection shuts down the furnace if temperature is outside of acceptable range (refer to controller's manual) or when the thermocouple is broken or malfunctions.
2.       Power failure protection resumes furnace operation right after the point of failure when power is reestablished.

Warning! NEVER fill furnace with explosive gases, including Hydrogen, Carbon Monoxide and Methane.
    General specifications
     Electrical requirements 110-240VAC, 50/60 Hz, single phase
     Tube material Quartz
     Maximum/constant working temperature 1200°C / 1100°C (1000°C constant temperature under vacuum)

     
     Rated vacuum pressure 0.001 Pa / 0.0075 millitorr / 0.0000 mbar
     Rated positive pressure 0.02 MPa / 150 torr / 3 psi
     Heating element type

     
      (Sweden) resistance coil wire (except for 50mm OD model)

     
     Thermo couple type K
     Refractory lining Mitsubishi (Japan) high quality 1500 grade fiber alumina
     Temperature controller Shimaden fp93 (Japan) with 4 programs and 
     40 segments (ie. 4 x 10 segments or 2 x 20 segments)
     Maximum heating & cooling rate   < 30°C / min
     Heating vs constant zone length 15.5" (8" for 50mm OD model) / 8" (4" for 50mm OD model)
     Temperature controller precision

     
     +/- 1°C
     Vacuum sealing flange kit Stainless steel vacuum sealing flanges with one vacuum gauge,
    two valves and four thermal ceramic blocks.      
     CE compliance

     
     Yes

     
     
    Size comparison
     Tube size (ODxlength)

     
     50mm x 700mm 60mm x 1m

     
     80mm x 1m

     
     100mm x 1m

     
     Maximum output 1.5 kW (110V) 2.5 kW 3.5 kW 3.5 kW
     Unit dimensions (WxDxH) 16x15x30" 21x18x43" 21x18x43" 22x20x45"
     Shipping dimensions (LxWxH) 45x20x35" 50x25x40" 50x25x40" 52x25x40"
     Shipping weight (Lb) 250 400 430 470
     
    Standard Package
    Lab Single-Zone 1200c Vertical Tube Furnace
    Lab Single-Zone 1200c Vertical Tube Furnace
    Lab Single-Zone 1200c Vertical Tube Furnace
    Lab Single-Zone 1200c Vertical Tube Furnace
    Lab Single-Zone 1200c Vertical Tube Furnace
    Lab Single-Zone 1200c Vertical Tube Furnace
    Lab Single-Zone 1200c Vertical Tube Furnace


    Lab Single-Zone 1200c Vertical Tube Furnace

    1 Standard exported package: Internal anticollision protection, external export wooden box packaging.
    2 Shipping by express, by air, by sea according to customers' requirements to find the most suitable way.
    3 Responsible for the damage during the shipping process, will change the damage part for you for free.

    Lab Single-Zone 1200c Vertical Tube Furnace