ICP-6810 Inductively Coupled Plasma Optical Emission Spectrometer
Application Field
Inductively coupled plasma emission spectrometer(ICP6800) developed by Macylab Instruments with many years of technology accumulation. It is used to measure traces and traces of various substances (dissolvable in hydrochloric acid, nitric acid, hydrofluoric acid, etc.),and the content of metal elements or non-metallic elements. It is stable and reliable and with high degree of automation and easy operation.
At present, instruments are widely applied in various fields such as rare earth, geology, metallurgy, chemical industry, environmental protection, clinical medicine, petroleum products, semiconductor, food, biological samples, criminal science, agricultural research and so on.
Application Environment
CONTENT RANGE
Shipping/Storage Ambient Temperature 15ºC-25ºC
Shipping/Storage Relative Humidity ≤70%
Atmospheric pressure 86-106 kPa
Power Supply 220±10V 50-60MHz
Working Humidity ≤70%
Working Temperature 15ºC-30ºC
RFPower technical parameter:
Circuit type: solid-state RF power supply, with function of automatch
Frequency: 27.12MHz±0.05%
Frequency Stability: <0.1%
Power Output: 800W-1200W
Power Output Stability: <0.3%
Escaped RF radiation: 30cm away from the instrument, electric field: E<2V/m
Sampling System Technical Parameter:
Output working coil inner diameter :25mm
Torque tube: Three concentric, external diameter 20mm
Coaxial nebulizer:Outer diameter 6mm
Double barrel atomizing chamber: Outer diameter 34mm
Gas Flow Controls
1. Plasma Argon Flowmeter:(100-1000)L/h (1.6-16L/min)
2. Auxiliary Argon Flowmeter:(10-100)L/h (0.16-1.66L/min)
3. Carrier Argon Flowmeter(10-100)L/h (0.16-1.66L/min)
4. Pressure Maintaining Valve(0-0.4MPa)
5. Cooling Water:Temperature: 20-25ºC, Rate of Flow>5L/min, Hydraulic Pressure>0.1MPa
Spectrometer
Optics:Czerny-Turner type
Focal length: 1000 mm
Grating: Ion Beam Etching Holographic Grating, 3600L/mm or 2400L/mm
Reciprocal linear dispersion: 0.26nm/mm
Resolution:≤ 0.007nm (3600 line grating);≤0.015nm (2400line grating)
Element | WL(nm) | Limit | Element | WL(nm) | Limit | |
La | 408.672 | <3.0 | Cr | 267.716 | <5.0 | |
Ce | 413.765 | <5.0 | Al | 396.152 | <5.0 | |
Pr | 414.311 | <5.0 | Zr | 343.823 | <5.0 | |
Nd | 401.225 | <5.0 | Ag | 328.068 | <3.0 | |
Sm | 360.946 | <10.0 | Sr | 407.771 | <1.0 | |
Eu | 381.967 | <1.0 | Au | 242.795 | <5.0 | |
Gd | 342.247 | <10.0 | Pt | 265.945 | <5.0 | |
Tb | 350.917 | <3.0 | Pd | 340.458 | <5.0 | |
Dy | 353.170 | <3.0 | Ir | 224.268 | <10.0 | |
Ho | 345.600 | <3.0 | Rh | 343.489 | <10.0 | |
Er | 337.271 | <3.0 | Ru | 240.272 | <5.0 | |
Tm | 313.126 | <3.0 | Ba | 455.403 | <1.0 | |
Yb | 369.419 | <1.0 | As | 228.812 | ≤15 | |
Lu | 261.541 | <3.0 | Sb | 206.833 | ≤15 | |
Y | 371.030 | <1.0 | Bi | 223.061 | ≤10 | |
Sc | 335.373 | <1.0 | Hg | 253.652 | ≤15 | |
Ta | 226.230 | <5.0 | Pb | 220.353 | ≤15 | |
Nb | 313.340 | <5.0 | Ga | 294.364 | ≤10 | |
Mn | 257.610 | <3.0 | Se | 203.985 | ≤10 | |
Mg | 279.553 | <1.0 | Sn | 242.949 | ≤20 | |
B | 249.773 | <10.0 | Te | 214.281 | ≤10 | |
Zn | 213.856 | <3.0 | Ta | 226.230 | ≤5.0 | |
Co | 228.616 | <3.0 | Th | 283.730 | ≤10 | |
Si | 251.611 | <10.0 | Tl | 276.787 | ≤30 | |
Ni | 232.003 | <5.0 | Re | 227.525 | ≤5 | |
Cd | 226.502 | <3.0 | Ge | 209.426 | ≤15 | |
Fe | 239.562 | <3.0 | Os | 225.585 | ≤1 | |
Ca | 393.366 | <1.0 | W | 207.911 | ≤10 | |
Mo | 281.615 | <5.0 | Cu | 324.754 | <3.0 | |
V | 310.230 | <5.0 | Li | 670.784 | ≤3 | |
Be | 313.041 | <1.0 | Na | 588.995 | ≤20 | |
Ti | 334.941 | <3.0 | K | 766.490 | ≤60 |
Wavelength range: 3600 line grating:(190nm~500) nm;2400 line grating: (190nm~800) nm
Minimum pace of stepping motor: ≤0.0006 nm
Exit Slit:12μm;Entrance Slit:10μm
Photoelectric Converter Performance
Photomultiplier tube specification: R293/R928
Negative HV on PMT: 0-1000V;Stability: <0.05%
Overall Performance
Wavelength range of scanning: 195nm~500nm(3600L/mm line grating);195nm~800nm(2400L/mm line grating)
Repeatability: RSD(Relative Standard Deviation)≤1.5%
Stability: RSD(Relative Standard Deviation) ≤2%
Detection Limits (µg/L):