China Quv Accelerated Weathering Tester UV Aging Testing Chamber, Find details about China UV Aging Chamber, UV Weathering Test Chamber from Quv Accelerated Weathering Tester UV Aging Testing Chamber
Studio size | 1140*400 *380mm(L*D*H) |
Overall size | 1300*500 *1460mm(L*D*H) |
Center distance of the lamp | 70mm |
Distance of specimen from lamp surface | The distance between the specimen and the nearest parallel surface of the lamp surface is about 50mm |
Wavelength range | UV-A wavelength range of 315 to 400nm |
Radiation intensity | within 1W/m2/340nm |
Temperature resolution | 0.1ºC |
Light temperature range | 50ºC~70ºC / temperature tolerance is ±3ºC |
Condensation temperature range | 40ºC~60ºC / temperature tolerance of ±3ºC |
Measuring range of blackboard thermometer | 30~80ºC / tolerance is ±1ºC |
Temperature control mode | PID self - tuning temperature control mode |
Humidity range | about 45%~70% R.H. |
Water tank requirements | water depth not more than 50mm, and with an automatic water supply controller |
Standard specimen size | 75×150mm 48pcs |
Recommended operating environment | 5~35ºC, 40%~85%R•H, 300mm away from the wall |
Voltage | 220V±5%, single-phase three-wire, 50Hz, 10A, required 15A leakage switch |