NIBP monitoring | Measurement method: Oscillometry Measurement mode: manual/auto. Auto measurement interval: adjustable from 1 to 255 minutes. Resolution: 1mmHg. Accuracy: Maximum Mean deviation±5mmHg, Maximum Standard deviation≤8mmHg. Self-check function. Overpressure protection: double protection for software and hardware. Measurement range: 10mmHg~270mmHg. |
SpO2 monitoring | Continuous real-time measuring. Adopt digital SpO2 technology, which has strong anti-interference capability. Real-time display SpO2, pulse rate, SpO2 Plethysmogram, bar graph, SpO2 and Pulse rate (PR) trend graph. |
Trend graph with the function of adaptive scaling | |
1.SpO2 Measuring | 1.Range: 0%~100% (Resolution: 1%) 2.Accuracy: ±2% in stage of 70%~100%, and unspecified in stage of less than 70% |
Pulse Rate Measuring | 1.Range: 30bpm~250bpm(Resolution: 1bpm) 2.Accuracy: ±2bpm or ±2%, whichever is greater |
Measurement Performance in Weak Filling Condition | SpO2 and pulse rate can be shown correctly when pulse-filling ratio is 0.4%, SpO2 error is ±4%; the Pulse rate error is ±2bpm or ±2%, whichever is greater |
Resistance to surrounding light | The deviation between the value measured in the condition of man-made light or indoor natural light and that of darkroom is less than ±1% |
Operational environment | |
Temperature | 5 ˚C~40 ˚C |
Relative humidity | 30%~75% |
Atmospheric pressure | 700hPa~1060hPa |
Specific EMC, climate, mechanical environment: Do not use a mobile phone in the vicinity of the veterinary monitor for the strong radiation field produced by mobile phone would interfere with the normal function of the veterinary monitor use. |