High frequency Xray equipmentsHigh Frequency 50kW, 150kV Radiography System
Mode:MED-X-HF50-R
(CE & FDA Marked)
High frequency desigh
Short Exposure time
High Kv (150kV), high mA (630 mA)
Small size focus (0.6mm)
AEC + APR function
LCD displays all radiographic modes
Friendly user interface and easy menu design.
Normal radiography, bucky radiography,
oblique radiography,
Ion chamber AEC mode can be selected.
Automatic heat capacity protection
New-designed table
Multi-style console installation, pedestal of wall handle
Modularized design, flexible intergration.
Performance feature:
1. Excellent High-frequency Generator
High-frequency generator output is stable and less scattering X-ray, the image fog degrees low, it
make the effective X-ray utilization increased, which saves the tube, but also effectively reduce the patient
radiation dose.
Transient exposure to eliminate the motion artifacts, it can accomplish the diagnostic in respiration, somebody
can't independently controlling breathing or crying children.
APR (Apparatus program radiography), enter as many as 600 set exposure conditions to program (5 kinds of
body type, 30 set positions, 4 set directions)
2. Humanization Design Console
LCD display module and surface lamination technology to ensure stable and reliable, unitized menu choices be
clear at a glance.
There are varieties of console installation, such as Wall-handle, pedestal, desktop placed. It could reduceoperating fatigue by human engineering design.
3. Solid Radiography table
The wide range of movement of Stand and radiography table to ensure patient to be covered the whole body by
X-ray tube, To fully meet the lateral position oblique position and the head of table and requirement of special
position.
Stand rotation makes stretcher patients to do diagnostic without moving.
Standard Configuration:
H.T. generator 50kW
H.T. cable 75kV
X-ray tube assembly Toshiba E784X
Radiographic table
Collimator
Option:
Vertical bucky stand
Ion Chamber
Standard Configuration:
No. | Component | Part Number | Quantity |
1 | High Frequency Generator | GFS501-1 | 1 |
2 | X-ray Tube Assembly | E7843X | 1 |
3 | HV cable | DL7512-1B | 1 |
4 | Collimator | XS1-2 | 1 |
5 | Radiography Table | SC3-3 | 1 |
6 | Source Assembly Stand | ZZ-5 | 1 |
Optional Configurations:
No. | Component | Part Number | Quantity |
7 | Vertical Bucky Stand | LS-3 | 1 |
8 | Ionization Chamber | SSMC610 3 fields | 1 |
Standard configuration Component Specifications
1.
High Frequency Generator Part Number: GFS501-1
Power: 50kW
Constant Frequency of Inverter: 25kHz
KV Range: 40~150kV
mA Range: 25~630mA
T ime Range: 2ms~5s
mAs 0.5mAs~500mAs
AEC function, APR
Power Supply: 380VAC, three-phase
Fault Detection and Diagnosis
Console: LCD Displays All Radiographic Modes
APR:600 positions
High Frequency Generator
2.
X-ray Tube Assembly Part Number: Toshiba E7843X
Focal Spot: 0.6mm/ 1.2mm
Power: 22/ 50kW
Anode Heat Capacity: 150kHU Target Angle: 12°
4.
Collimator Part Number: XS1-2
External Dimension: 250mm×224mm×250mm
Radiation Field(SID=65cm) Max: 35cm×35cm; Min: 0cm×0cm
Operation Mode: Manual
5.
Radiography Table Part Number: SC3-3
External Dimension(Length* W idth*Height) : 2100mm* 800mm*650mm
Tabletop Movement(Longitudinal):900mm
Tabletop movement(Lateral): 240mm
Tabletop Locking Mode: Electromagnetic lock braking force 100N
Tabletop Locking Control footswitch (normally closed) Bucky V ibration Mode Fixed
Bucky specification f0: 100cm, r: 10, N: 40L/cm
Distance between Tabletop and Bucky Film 65mm
Collimator Longitudinal travel 400mm
Bucky locking: Electromagnetic, braking force 50N
Radiography table lateral moves to limit position, evenly on the surface of the bed to
Load of 100 kg. and the table could run normally, its maximum weight 135 kg, conform to the
requirements of relevant national standard
6.
Source Assembly Stand Part Number: ZZ-5
X-ray Tube components Up-and Down Motion Range: 1220mm
Movement Range Between Tube Focus and Film(SID) : 1100mm
X-ray Tube Assembly Longitudinal Motion Range:2380mm
X-Ray Radiation Component Rotating Range
Around Support Arm Axis: 120° 120°
X-Ray Radiation Component Rotating Range
Around the Vertical Axis of the Stand :0°±90° and 90° mechanical positioning lock
X-Ray Tube Assembly Locking Force: Electromagnetic locking branking force >= 100N
Automatic measuring the tube rotation angle and SID, those could be showed by digital display.
7.
Vertical Bucky Stand Part Number: LS-3
Vertical Movement: >= 1000mm
Maximum Dimension: 8" x 10", 14" x 17"
Grid(replaceable)
N: 40L/cm; r8 r10 r12; f0: 150cm 180cm 200cm
8.
Ionization Chamber Part Number: Claymount SSMC610
AEC function, 3 fields
Exposure dose range 1 100uGy
Exposure time range:1ms 5s
Welcome to visit us for High frequency Xray equipments!