Bojin Medical Surgical Power Tool Reciprocating Saw Bj4309

China Bojin Medical Surgical Power Tool Reciprocating Saw Bj4309, Find details about China Surgical Power Tool, Medical Saw from Bojin Medical Surgical Power Tool Reciprocating Saw Bj4309

Model NO.
BJ4309
Item
Bj4309
Name
Reciprocating Saw
Oscillations
0-15000
Battery
Rechargeable
Use for
Trauma Operation
Packing
Carton,Aluminum Box
Warranty
2years
Certificate
Ce,ISO,FDA
Color
Silver
Trademark
bojin
Transport Package
Carton, Aluminun Box
Specification
10*40*40
Origin
Shanghai
HS Code
9018909010
Model NO.
BJ4309
Item
Bj4309
Name
Reciprocating Saw
Oscillations
0-15000
Battery
Rechargeable
Use for
Trauma Operation
Packing
Carton,Aluminum Box
Warranty
2years
Certificate
Ce,ISO,FDA
Color
Silver
Trademark
bojin
Transport Package
Carton, Aluminun Box
Specification
10*40*40
Origin
Shanghai
HS Code
9018909010
Bojin Medical Surgical Power Tool Reciprocating Saw Bj4309
Bojin Medical Surgical Power Tool Reciprocating Saw Bj4309
Bojin Medical Surgical Power Tool Reciprocating Saw Bj4309

Specification:
Output power≥90w
Speed(r.p.m) /
Oscillations (osc/min)0-15000
Torque(Nm)/
Operation(V) voltage12
Sterilization155°C
Swings (°) /
Noise(db)≤70

Notes:
1.Sterilize temperature ≤ 135 °C (Improved type ≤ 155 °C ),if the 
    temperature is over 135 ° C ,the motor will be damaged.
2.Normal bone drill can only hold drill bits less than 6mm,acetabulum 
    reaming drill with big torque,rpm is around 300,canulate drill is used 
    with nail or K wire ,rpm is around 600.
3.The battery need to be get out of the pwer tool after surgery. Otherwise 
    the small automatic discharge of the electric will make the life of the 
    battery short.
4.The battery cannot be sterilized ,it is not autoclavable.
5.The cranial bur must be washed right after the surgery by purger in 
    order to aviod the blood stick in the bur,then we should make it dry.
6.The battery space in the handpiece of the tool should not have any 
    water,after sterilization.
7.The cranial bur blade is made of medical stainless steel,in order to 
    avoid broken,so it is flexible and not very hard,it is easy to be attrite, 
    we suggest each blade only use for 1 surgery.
Bojin Medical Surgical Power Tool Reciprocating Saw Bj4309
Bojin Medical Surgical Power Tool Reciprocating Saw Bj4309