Hv-300e Electrosurgical Generator

China Hv-300e Electrosurgical Generator, Find details about China Electrosurgical Generator, Generator from Hv-300e Electrosurgical Generator

Model NO.
HV-300E
Model NO.
HV-300E
HV-300E Electrosurgical Generator
Features:
Intelligent Device System 
The intuitive design and user-friendly set-up of the Electrosurgical Generator (diathermy) for us in the modern operating room, it features both Monopolar and bipolar functions to satisfy all the surgeries demands with safety, flexibility, reliability and convenience.
REM (Return Electrode Monitoring): 
Return electrode contact quality monitoring system (REM). The REM system continually monitors patient impedance levels and deactivates the generator if a fault in the patient/return electrode contact is detected, which minimize the danger of burning incidents.
Compatible with Other Equipments:
▪ Optimum Smoke Evacuation System
Multilanguage Available:
Language options: English, Spanish, Portuguese, Turkish and etc
Application of Use: 
General Surgery; Dermatology; Vascular Surgery; Obstetrics and Gynecology
ORL/ENT; Minimally Invasive Surgery (MSI)
Cerebral surgery;Neurosurgery, Orthopedics & Plastic Surgery.
Low Frequency Leakage:
Normal Polarity, Intact Chassis Ground: <10μA
Normal Polarity, Ground open< 50μA
Reverse Polarity, Ground Open< 50μA
Sink Current, 140V Applied, All inputs < 50μA
Power Readouts
Agree with actual power into rated to within
±20% or 5W, whichever is greater, Dosage
Error monitor.
Hv-300e Electrosurgical GeneratorHv-300e Electrosurgical GeneratorHv-300e Electrosurgical GeneratorHv-300e Electrosurgical Generator
Technical Specification:
Main Frequency512 KHz
Power Supply110V/60Hz;220V/50Hz
Weight6.5KG
Dimension28*38*40
High Frequency LeakageMonopolar: < 150mA;
Bipolar < 60mA
Low Frequency LeakageNormal Polarity, Intact Chassis Ground: <10μA
Normal Polarity, Ground open< 50μA
Reverse Polarity, Ground Open< 50μA
Sink Current, 140V Applied, All inputs < 50μA
Power ReadoutsAgree with actual power into rated to within ±20% or 5W, whichever is greater, Dosage Error monitor.
Hv-300e Electrosurgical Generator
Hv-300e Electrosurgical Generator
Hv-300e Electrosurgical Generator