Tealth Detachable 20: 1 Implant Surgery Dental Contra Angle Handpiece

China Tealth Detachable 20: 1 Implant Surgery Dental Contra Angle Handpiece, Find details about China Dental, Dental Handpiece from Tealth Detachable 20: 1 Implant Surgery Dental Contra Angle Handpiece

Model NO.
R20N-04C
Application
Adult
Loading Way
Gland Type
Head Diameter
Middle Head
Exhaust Way
Exhaust Outside
Number of Holes
E-Type
Connection
Without Quick Connector
Rotating Speed
Low-Speed Handpiece
Gear Ratio
20:1 Reduction
Noise
Less Than 70dB
Max Torque
70n.Cm
Trademark
TEALTH
Transport Package
Plastic Box
Specification
Aluminum, stainless steel
Origin
China
HS Code
90184100
Model NO.
R20N-04C
Application
Adult
Loading Way
Gland Type
Head Diameter
Middle Head
Exhaust Way
Exhaust Outside
Number of Holes
E-Type
Connection
Without Quick Connector
Rotating Speed
Low-Speed Handpiece
Gear Ratio
20:1 Reduction
Noise
Less Than 70dB
Max Torque
70n.Cm
Trademark
TEALTH
Transport Package
Plastic Box
Specification
Aluminum, stainless steel
Origin
China
HS Code
90184100
Name: Detachable 20:1 implant contra angle dental handpiece
Order code: R20N-04C
Tealth Detachable 20: 1 Implant Surgery Dental Contra Angle HandpieceTealth Detachable 20: 1 Implant Surgery Dental Contra Angle Handpiece

Tealth Detachable 20: 1 Implant Surgery Dental Contra Angle Handpiece
Pack list :
1.20:1 contra angle  *1 pcs
2.Cap spanner * 1 pcs
3.Operation manual  * 1 pcs
4.E -type Oil spray nozzle *1 pcs
5.Internal water tube * 1 pcs
6.Tube fixture holder *1pcs

Parameter:
Push button type, Fiber optic
Torque force: up to 70 N.cm 
Rotation speed :5~2000RPM
Ergonomics coating ,more  hygeian.
Sterilization : up to 135 ºC
Applicable bur : Φ2.345-2.355mm
Fitting with E type motor , can be avaible to match up with Swiss, Japan, Korea famous brand implant system.


 Insert or remove bur or drill 
1) Inserting(Fig.1,Fig2)
Press the push button firmly, insert bur and turn the rod in clockwise direction, till the bur to the end.
Tealth Detachable 20: 1 Implant Surgery Dental Contra Angle HandpieceRemoval
Press the push button firmly and remove the bur.(Fig.3)
Tealth Detachable 20: 1 Implant Surgery Dental Contra Angle Handpiece
Handpiece inserting & removal:
Inserting
Insert the handpiece into the motor and turning all the way until it click to stop.(Fig.4)
Tealth Detachable 20: 1 Implant Surgery Dental Contra Angle Handpiece
2) Removal
Grip both handpiece and motor respectively,pulling and separating.

Notes:
Please do not insert or remove handpiece before motor complete stop turning.
Compatible with dental motor implant system in accordance with ISO standard.
Compatible with the electro-motor .
The structure of connector conform with ISO3964 standard.

Water Spray: 
1) External water spray
    Connect the water pipe and the external water spray pipe of handpiece.(Fig.5)
Tealth Detachable 20: 1 Implant Surgery Dental Contra Angle Handpiece
2) Internal water spray
    Use internal bur to operate. Water pipe connect to the internal water spray pipe, and make the internal water spray pipe fixed with the pipe fixture. (Fig.6)
Tealth Detachable 20: 1 Implant Surgery Dental Contra Angle Handpiece
3)  Mixed water spray
    Use internal bur to operate, and follow step 1 and 2.

 Maintenance & Care:
1. Cleaning
    Remove bur from handpiece after using and brush. soak the head into clean water and operate for 15 seconds(Fig.7). Then autoclave handpiece.
Tealth Detachable 20: 1 Implant Surgery Dental Contra Angle Handpiece
    If the handpiece is the type of head can remove one, please  take off the head and wash it and the body with clean water.   
2. lubricating(Fig.8)
Lubricate oil with E type oil nipple insert the handpiece and spray 2 or 3 seconds.
Tealth Detachable 20: 1 Implant Surgery Dental Contra Angle Handpiece
    If handpiece head dirty enough, please remove the back cap with spanner and spray oil inside(Fig.9 and 10).
 Tealth Detachable 20: 1 Implant Surgery Dental Contra Angle Handpiece

3. Sterilization
    It shall be sterilized before handpiece is used at the first time: Autoclave the handpiece for 12 min. at 135ºC
After each treatment, sterilizing handpiece in autoclave according to the following steps:
    1)Remove the bur and brush clean handpiece, clean outside of handpiece with an alcohol solution.
    2)Lubricate handpiece.
    3)Place into autoclaving pouch and mark.
    4)Autoclave the handpiece for 12 min. at 135ºC.

Note:
Please do not use the acid solution to scrub, inject or soak the handpiece.
Please do not autoclave the handpiece with residual chemical substances.
Handpiece should be packed well and place at clean and dry room without dust, sulfur and salt.
Please do not touch the handpiece immediately after autoclaving.