14C Urea Breath Test Helicobacter Pylori Detector with HP Test Kit
CHL-HP14C20/P


Applications of C14 Urea Breath Test:
Dyspepsia initial diagnosis, post treatment follow-up of H.Pylori infection;
The patients who suffers Gastric Ulcer, duodenum ulcer, chronic active gastritis, antral gastritis, Gastric mucosa, lymphoid tissue malignant tumor disease;
Precaution of gastic cancer or the ones with family history of gastric cancer;
Refuser of gastroscopy;
Long term user of NSAID(non steroidal anti-inflammatory drugs);
Idiopathic thrombocytopenic purpura(ITP);
Patients already  take endoscopy;
Physical examinations;
Patient wishes

14c Urea Breath Test Helicobacter Pylori Detector with HP Test Kit
The urea breath test is used to diagnose H. Pylori infection.
 
UBT can be widely used in digestive and gastroenterology, physical examination centers,nuclear medicine hospital. 
 
Highlights: non-invasive, accurate, fast, convenient.
 
H. Pylori Test Analyzer CHL-HP14C20A is designed to provide the user with a reliable and advanced intelligent device that is both easy to use and meets the strict clinical demands for Helicobacter pylori diagnosis.
Helicobacter pylori (H pylori) is the primary cause of gastritis and peptic ulcer disease and has been linked to the onset of gastric cancer in some individuals. The need for a simple, non-invasive and accurate diagnostic technology for the detection of H. Pylori infection was met by the development of the urea breath test, which allows infection status to be evaluated without the need for costly and invasive endoscopies.
 
Advantages: 
Accurate: Sensitivity 95%, specificity 95%-100%.
Comfortable: Non-invasive, no endoscopic tube insertion, well tolerated by patients.
Convenient: Near-patient testing, the analysis made on-site, minimal inconvenience for patients.
Simple: Easy operation, no special handling required.
Swift: The entire test is finished in 30 minutes, test result available on site.
 
Product Highlights:
High accuracy with DPM results
Streamlined design, elegant appearance
PC connection, professional software
Automatic measurement and patient data storage
 
Specifications:
1. Background Count Rate ≤ 50 CPM
2.14C Detection Efficiency≥ 15%
3. Measuring Time: 250 seconds
4. Power ≤ 20W
5. Dimension: 400mm× 260mm× 200mm
6. Weight: 12Kg
14c Urea Breath Test Helicobacter Pylori Detector with HP Test Kit

Warm-up after power-on
Install the instrument in a suitable working environment and connect to power. Turn on the power switch of the instrument. When it reaches a stable state after warming up for 30 minutes, it can be normally used.
 
Calibration
a. Insert the attached standard background card into the measuring slot of the instrument to calibrate the background for 1000 seconds without inserting a card. The calibration result shall be BGC ≤ BGM. Insert the attached standard background card into the measuring slot for sample measurement (set the parameter MOD_CON to 0), and the measurement result C should be negative.
 
b. Insert the attached standard efficiency card into the measuring slot for sample measurement (set the parameter MOD_CON to 0), and the measurement result C should be within ±30% of the labeled value.
 
When the conditions of a and b are met simultaneously, the instrument can be operated normally.
 
Sample collection:
a. The subject takes one 14C-Urea on an empty stomach or two hours after eating.
 
b. Sit for 15-25 minutes.
 
c. Unpack the collection card and connect the mouthpiece to the card body. The subject blows into the card through the mouthpiece with appropriate force. The subject may exchange breath during blowing, but DO NOT draw in a breath! Stop blowing when the card indicator turns from orange to yellow (about 1-3 minutes) or after 3 minutes even though the color is not completely changed.
 
d. Fill in the information of the subject on the collection card after sample breath is collected.
 
Sample testing:
Insert the sample card into a measuring slot, and the test will start automatically. The test result will be displayed on the screen when the test completes.
 
Result printing:
The test reports can be printed either from an external printer or built-in micro printer.

14c Urea Breath Test Helicobacter Pylori Detector with HP Test Kit