China Clinic ICU Multipara Patient Monitor, Find details about China Patient Monitor, Clinic ICU Multipara Patient Monitor from Clinic ICU Multipara Patient Monitor
ECG | Lead mode: 3-lead or 5-lead Lead selection: I, II, III, aVR, aVL, aVF, V Waveform: 5-lead, 2-channel 3-lead, 1-channel Gain: 2.5mm/mV, 5.0mm/mV, 10mm/mV, 20mm/mV, 40mm/mV Scan speed: 12.5mm/s, 25 mm/s, 50 mm/s |
HR: | Measurement and alarm range: 15~350bpm Accuracy: ±1% or ±1bpm, whichever is greater Alarm accuracy: ± 2 bpm Resolution: 1 bpm ST-segment monitoring: Measurement and alarm range: -2.0mV~+2.0mV Accuracy: -0.8mv~+0.8mv ±0.04mv or ±10%, whichever is greater Other range: unspecified Arrhythmia analysis: ASYSTOLE, VFIB/VTAC, COUPLET, BIGEMINY, TRIGEMINY, R ON T, VT>2, PVC, TACHY, BRADY, MISSED BEATS, PNP, PNC Pacemaker: yes |
RESPIRATION | Method: R-F(RA-LL) Impedance Respiration rate: Measurement and alarm range: 0~150rpm Resolution: 1 rpm Measurement accuracy: ±2 rpm Alarm accuracy: ±3 rpm Apnea alarm: 10~40s Scan speed: 6.25 mm/s, 12.5 mm/s, 25 mm/s |
NIBP | Method: Oscillometry Mode: Manual/AUTO/Continuous Measurement interval in AUTO mode: 1/2/3/4/5/10/15/30/60/90/120/240/480/960 minutes Measurement period in Continuous mode: 5 minutes Measurement and alarm range: 10 ~ 270 mmHg Resolution: 1 mmHg Cuff pressure accuracy: ±3 mmHg Measurement accuracy: Maximal mean deviation: ±5 mmHg Maximal standard deviation: 8mmHg Over-pressure protection: Adult mode: 297±3 mmHg Pediatric mode: 240±3 mmHg Neonatal mode: 147±3 mmHg |
SpO2 | Measurement and alarm range: 0 ~ 100% Resolution: 1% Measurement accuracy: 70%~100%: ±2%; 0%~69%: unspecified |
PR | Measurement and alarm range: 30 ~ 250 bpm Measurement accuracy: ±2 bpm or ±2%, whichever is greater |
TEMP | Channel: dual-channel Measurement and alarm range: 0 ~ 50ºC Resolution: 0.1 ºC Accuracy: ±0.1 ºC |
EtCO2 | Method: Sidestream or Mainstream Measurement range for CO2: 0~150mmHg Resolution for CO2: 0 ~ 69 mmHg: 0.1 mmHg 70 ~ 150 mmHg: 0.25 mmHg Accuracy for CO2: 0 ~ 40 mmHg: ±2 mmHg 41 ~ 70 mmHg: ±5% 71 ~ 100 mmHg: ±8% 101 ~ 150 mmHg: ±10% AwRR range: 2~150 rpm AwRR accuracy: ±1 rpm Apnea alarm: yes |
IBP | Channel: dual-channel Label: ART, PA, CVP, RAP, LAP, ICP, P1, P2 Measurement and alarm range: -10~300 mmHg Resolution: 1 mmHg Accuracy: ±2% or 1mmHg, whichever is greater Power supply:AC 100-240V, 50/60Hz Safety classification: Class I, type CF defibrillation-proof applied part |