China Infant Phototherapy Incubator Warmer System (baby incubator) Medical Equipment, Find details about China Medical Baby Incubator, Hospital Baby Incubator from Infant Phototherapy Incubator Warmer System (baby incubator) Medical Equipment
Infant Incubator
Medical Device New Born Baby Use Incubator(One-Stop Service)
Name | Infant Incubator(Medical Infant Incubator) |
Feature 1 | Air temperature display and servo contro |
Feature 2 | 24 hours Parametric trend diagram display |
Feature 3 | Double skin temperature sensor |
Certification | CE,ISO13485 |
Payment | L/C,T/T,Cash |
Package | Each unit packed in 3cartons with Wooden |
Delivery | 25 days |
Product feature:
-Servo Control Mode: air temperature, skintemperature, display humidity;
-Parameter Displaying: set temperature,air temperature, skin
temperature, internal humidity, heat power rate;
-Audible & Visual Alarms: powerfailure, sensor failure,
-higher & lower temp, air & skin temperature override,
fan failure, system failure;
->37ºC Preset Mode: power failure reset memory, alarm silence,
-keypad lock, self-testfunction;
-Hood with dual slope; 6 hand-ports;
-RS-232 Output Connector (use for networksystem);
-IV poles;
-Protective guard rail;
-Storage cabinets;
-Can opiton humidity control.
Specifications of hood
Mattress to Hood: 450 mm
Observation side of hood: Two incline
Operate port: 6 (2 Iris ports)
Tubing grommets: 4
Front access panel: 826×310 mm
Mattress Tray Size: 630×360 mm
Infant bed tilt adjusting ±10° step-less
Technical Specifications
Power requirement ≤400VA
Temperature control mode:
Air and skin servo control
Air Temp Control Range: 20.0ºC-37.0ºC
Temperature override mode to: 37.1ºC-39.0ºC
Skin Temp Control Range: 20.0ºC-37.0ºC
Temperature override mode to: 37.1ºC-38.0ºC
Skin temp sensor accuracy: ≤0.3ºC
Humidity display range: 5%-99%
Temperature fluctuation: ≤0.5ºC
Mattress temperature uniformity: ≤0.8ºC
Temperature rise time: ≤45 min
Noise level within hood environment: ≤55dB
Alarms
Power failure alarm
Fan failure alarm
Sensor failure alarm
High air temperature alarm: +3.0ºC
Low air temperature alarm: -3.0ºC
High skin temperature alarm: +1.0ºC
Low skin temperature alarm: -1.0ºC
Over air temperature alarm: ≤38.0ºC; ≤40.0ºC
Over skin temperature alarm: ≤38.0ºC; ≤39.0ºC
System failure alarm
Alarm silence function
Operational condition
Environment temperature: +20ºC-+30ºC
Environment relative humidity: 30%-75%
Environment air velocity of flow: <0.3m/s
Others
Mattress tray Max. load 5kg
Transfusion pole Max. load 2kg
Cabinet Max. load 6kg