Display | 12.1"color TFT Rolling and refreshing waveform display Resolution: 800×600 Multi display selectable including: Standard display Large-font display Trend coexist display OxyCRG dynamic view display 7 full ECG waveforms display Bed-to-bed view display(optional) |
Trace | 9 waveforms (7 ECG, 1 SPO2 and 1 RESP) Sweep speed:6.25mm/s, 12.5mm/s,25mm/s,50mm/s |
Indicator | Power/battery indicator light QRS beep and alarm sound |
Battery | Rechargeable lithium battery, 12v/4AH Max.24 hour for charging, 2 hours for continue working |
Trend | Parameter graphic and tabular trends: 5s/piece, 8 hours; 1min/piece,168 hours(24hour×7days) 5min/piece, 1000 hours. |
Storage | NIBP: 1000 groups Alarm: 200 groups SD card external storage (optional) |
Alarm | User-adjustable High, Medium and Low limits 3-level Audible and visual alarm |
Networking | Connected to central monitoring system TCP/IP netting platform Wifi connector (optional) |
Recorder | Build-in, thermal array Plethysmogram caveform:2 /3 channels Record mode: manual, on alarm, time-defined Recording width: 50mm Printing speed: 50mm/s |
Recording type | Frozen waveform record NIBP recall record Trend table record Alarm record Fixed -time record |
ECG |
Lead mode: 5 -lead(R, L, F, N, C) (3/12 lead ECG for optional) Lead selection: I, II, III, avR, avL, avF, V Waveform: 3 and 7 channel selectable Gain selection: 0.5mm/mv,1mm/mv,2mm/mv Sweep speed: 6.25mm/s;12.5mm/s;25mm/s;50mm/s Heart rate range: Adult: 15~300bpm; Neonate/pediatric: 15~350bpm Accuracy: +1bpm or +1%, which is greater Resolution: 1bpm Filter: surgery mode: 1~20Hz Monitor model: 0.5~40Hz Diagnostic mode: 0.05~130Hz Scaling signal: 1mv, +3% Protection: withstand 4000VAC/50 voltage isolation against electrosurgical interference and defibrillation Alarm range: 15~350bpm S-T segment detection: Measurement range: 2.0mV~+2.0mV Alarm Range: -2.0mV~ +2.0mV Accuracy: -0.8mV ~+0.8Mv Error: +0.02Mv Arrhythmia analysis: YES |
SPO2 |
Measurement range: 0~100% Wave speed: 12.5mm/s;25mm/s Resolution: 1% Accuracy: +2% (70-100%); 0-69% unspecified Alarm range 0~100% Lower perfusion situation |
Pulse Rate(PR) |
Range: 20~300bpm Resolution: 1bpm Error: +1bpm or +2%, whichever is greater |
NIBP |
Method: Digital Automatic oscillometric Operation mode : Manual/Automatic Auto measurement time: Adjustable (1~720min) (1,2,3,4,5,10,15,30,60,90,120,240,480,720min) Measurement Unit: mmHg/Kpa selectable Measurement types: Systolic, Diastolic, Mean Measurement Range: Range of Systolic pressure: Adult: 40~270mmHg Pediatric: 40~220mmHg Neonate: 40~135mmHg Range of mean pressure: Adult: 20~235mmHg Pediatric: 20~165mmHg Neonate: 20~110mmHg Range of diastolic pressure: Adult: 10~215mmHg Pediatric: 10~150mmHg Neonate: 10~100mmHg Over-pressure protection: Double safety protection Resolution: 1 mmHg Alarm: Systolic, Diastolic, and Mean |
RESPIRATION |
Method: RA-LL impedance Measurement range: Adult: 7~120rpm; Neonate/Pediatric: 7~150rpm Apnea alarm: YES, 10~40s Resolution: 1rpm Accuracy: +2rpm |
TEMPERATURE |
Compatible probe: YSI or CYF Measurement range: 5~50ºC Resolution: 0.1ºC Accuracy: +0.1ºC Refreshing time: about 1 Average measuring time: <10s |