Acrel Hospital Ground Insulation Fault Detection Device Ail150 -4 for Surgical Rooms

China Acrel Hospital Ground Insulation Fault Detection Device Ail150-4 for Surgical Rooms, Find details about China Medical Isolated System, Unearthed Systems from Acrel Hospital Ground Insulation Fault Detection Device Ail150-4 for Surgical Rooms

Model NO.
AIL150-4
Communicaiton
RS485 Modbus-RTU
Trademark
ACREL
Transport Package
Fob Shanghai
Origin
China
Model NO.
AIL150-4
Communicaiton
RS485 Modbus-RTU
Trademark
ACREL
Transport Package
Fob Shanghai
Origin
China
Product Description
ACLP10-24 DC regulated power supply for Medical Insulation Power System

Acrel Hospital Ground Insulation Fault Detection Device Ail150-4 for Surgical Rooms
General
Hospital isolated electrical power supply dc stabilized voltage supply
The medical hospital isolated power supply system is mainly used in important Class 2 medical locations such as operating rooms,ICU/CCU intensive care unit, providing safe, reliable and continuous power distribution for the important equipment at these locations.
Medical IT system insulation monitoring and fault locating products (five-piece set) include AITR series medical isolation transformer, AIM-M100 medical intelligent insulation monitoring instrument, AKH-0.66P26 current transformer, ACLP10-24 DC power supply, and AID series(AID100,AID120,AID130,AID150) centralized alarm and display instrument



Features
The power supply module is designed to supply AID series alarm and display instrument.
The output supply voltage is DC 24V
with a total power consumption of maximum 3 VA.
A maximum of 3 AID10/AID120/AID150 can be supplied.
The power supply module is designed to supply AID series alarm and display instrument.
The output supply voltage is DC 24V with a total power consumption of maximum 3 VA.
A maximum of 3 AID10/AID120/AID150 can be supplied.
Voltage stability
Small ripple
Din 35mm



Technical Parameters
 
Input voltageAC 220V(fluctuating range±10%)
Frequency50/60Hz
Power3W
Output voltageDC 24V±5%
Voltage regulation factor≤30%
Temperature rise≤20ºC
Dielectric intensity4000V AC/minute

Dimension(Unit:mm)

Acrel Hospital Ground Insulation Fault Detection Device Ail150-4 for Surgical Rooms


Wiring
Acrel Hospital Ground Insulation Fault Detection Device Ail150-4 for Surgical Rooms




Typical Connection
Acrel Hospital Ground Insulation Fault Detection Device Ail150-4 for Surgical Rooms



Quotation
Acrel Hospital Ground Insulation Fault Detection Device Ail150-4 for Surgical Rooms





Acrel Hospital Ground Insulation Fault Detection Device Ail150-4 for Surgical Rooms





Acrel Hospital Ground Insulation Fault Detection Device Ail150-4 for Surgical Rooms


Acrel Hospital Ground Insulation Fault Detection Device Ail150-4 for Surgical Rooms

Acrel Hospital Ground Insulation Fault Detection Device Ail150-4 for Surgical Rooms