China New Type Medical First Aid Aed Automated External Defibrillator Monitor, Find details about China Patient Peacemaker, Vital Sign Monitor from New Type Medical First Aid Aed Automated External Defibrillator Monitor
New product Automatic External Defibrillator
Features
Category | Specification |
Output | Biphasic Truncated exponential |
Energy Sequence(adult ) | 150/150/200(Default) 150/200/200 200/200/200 Note: the Energy sequence can be changed using the computer IR port. |
Energy Sequence(Pediatric) | 50/70/85(Default) 70/70/85 70/85/85 85/85/85 Note: the Energy sequence can be changed using the computer IR port. |
Charge Time | 10 sec. to 200 J |
Analysis Time | 9 sec. |
The maximum time from the initiation of rhythm analysis to readiness for maximum Energy discharge after 15 shocks. | Less than 30 seconds |
The maximum time from initially switching power on to readiness for maximum Energy discharge after 15 shocks.. | Less than 35seconds |
Audible Prompts | 19 audible prompts |
Visual Prompts | LED/LCD |
Controls | six buttons - On/Off, Shock, and 4 software-configurable buttons |
Output Energy Accuracy | ±15% into any impedance from 25 to 175Ω |
The Maximum Voltage | 1800±20V |
Output disabled when the PATIENT impedance is outside limits | 20Ω to 200Ω |
Recovery of the ECG input after defibrillation | <10s |