China Biphasic Aed Portable Defibrillator Monitor, Find details about China Automatic External Defibrillator Monitor, Defibrillator Monitor from Biphasic Aed Portable Defibrillator Monitor
Defi Xpress is a compact, durable, light weighted defibrillator which integrated monitoring, manual defibrillation, AED and pacer (option). It is a professional biphasic defibrillator-monitor suitable for hospitals and clinics.
Two Defibrillation Mode:AED/Manual
Advanced Biphasic Defibrillation Technology
Features
1:Defibrillator
2:Biphasic energy output: 2-360J
3:ECG monitoring
4:SpO2
5:AED
Display
Screen Type: Highresolution display
Screen Size. 7 inches(17.7 cm) diagonally
Sweep Speed: 25mm/sec
Information: Heart Rate. Lead/Pads, AlarmOn/Off, SpO2, AED Functions and Prompts.
Alarm Selection and Limits. Delivered Energy
Defibrillator
Waveform: Biphasic
Energy Display: Monitor display indicates both selected and delivered energy
Charge Prompt Type: Voice and visual prompts
Electrode Impedance Measurement Range: O- 250 ohms
ECG Monitoring
Patient Connection | 3-lead ECG cable, or 5-lead ECG cable, paddles |
Lead Selection | Displayed on monitor, paddles, I,I,Il,AEl,AVR, AVL, AVF, V |
ECG Size | 0.25, 0.5, 1, 15, 2, 4 cm/mV diplay on monitor. |
Heart Rate | 20- 300BPM |
Heart Rate Alarm | On/Off displayed on monitor, user- selectable |
Smart Alarms | Beeper/voice prompts indicate shockable rhythm |
Battary
Type: Rechargeable, NiMH battery. 12V
Operating Time: For a new, fully charged battery 60 defibrillator discharges at maximum energry, or 3 hours minimum of continuous
ECG monitoring.
Additional paremeters will effect operating time with different functions.
SpO2 Module
Measurement Range: 30~100%,±2% between 80%~90%,others ±5%
Alarm Range: User set high limit and low limit
Alarm Accuracy: ± 1O% within setting values
Alarm time Accuracy: less than 12 sec
Recorder
Paper: 50mm thermal
Speed: 12 5mm/sec, 25mm/sec, 50mm/sec.User-selectable 6-second delay
Printing Method: High-resolution thermal array print head
Printout Modes: Manual or automatic, user- -configurable
On/Off Control: Front panel and paddle
Automatic Function: 9- second recording initiated by alarm
activation or defibrillator charge or defibrillator discharge
AED Mode
AED Function: Auto analyze and charge x3 with programmable auto energy level selection,screen prompts, and voice prompts
Shockable Rhythms: Ventricular fibrillation
Ventricular tachycardia
QRS complex wave duration
Charge Control: Control on device front panel, press key on paddle
Manual Mode
Energy Selection: Selectable at2, 5,7, 10, 20, 30, 50, 70,100, 150, 200, 300, 360 joules
Synchronized Mode: Synchronizes defibrillator pulse to patient's R-wave.
"SYNC" message displayed on monitor