Vet Clinic Hospital Use Anesthesia Machine Veterinary 
 
10.4" screen displaysinformation area,alarm area, monitoring area,setting information
electronic flowmeter,instantly know the fresh gas flow to animals.
Real time pressure-time,flow-time loop oscillogram and high precision ETCO2,O2concentration detection function included
ETCO2 module and Anesthesia Gas Scavenging System (AGSS)are optional

Ventilation mode: IPPV, A/C,PCV,SIMV,SIGH,MANUAL
Vt:0,20-1500ml
I;E:4:1-1:8
Gas source:N2o and O2
Oscillogram:P-T,F-T,P-V loop,V-T,ETCO2-T

Gas supply

  1. Gas supply                                          O2(For medical use)
                                         N2O(For medical use)
  1. Input pressure range                         280 kPa~600 kPa
  2. Output pressure of regulator
(gas cylinder supply)                   less than 440 kPa
  1. Safety valve                                          750kPa
  1. Negative pressure air source            ≥75 L/min

 Flow rate

  1.  Adjust range of flow meter
O                                            0 L/min~10 L/min
N2O                                        0 L/min~10 L/min
  1.  precision(20ºC 101.3kPa,supply pressure 440 kPa)
The precision from 10 % of full scale to full scale is ±10 % of indicated value
The precision lower than 10 % of full scale is class 4.
  1. Linkage control of Oand N2O flow rate
Adjust O2 concentration to no less than 25 % according to proportion.
  1. Flow rate of oxygen flush valve                  25 L/min~ 75 L/min 

Power supply

  1. Voltage                                                 100-240 V ~
  2. Frequency                                            50/60 HZ
  3. Input power                                          80VA
  1. Fuse                                                      
--Main power                                T2AH 250V
--Internal power                       T4AH 250V
 

Ventilation mode

--IPPV
--A/C
--SIMV                                              
--PCV

Ventilation parameters

Specification

  1. Frequency
--Adjusting range                         1 bpm ~ 100 bpm
(Under SIMV:1 bpm~ 40 bpm
All mode except SIMV:
VTH   4 bpm~40 bpm
VTL   20bpm~100 bpm )
--Allowable error                                  ± 15 %
  1. Inhalation time(I:E)(Tinsp) 
--Adjusting range                               VTH   4:1(1:0.25) ~ 1:8
                                           VTL   2:1(1:0.5) ~ 1:8
Tinsp   0.2s ~ 12.0s
--Allowable error                            ± 15 % or ± 0.1 s,
whichever is the greater
  1. Tidal volume(VT)
--Adjusting range                               0,20 mL ~ 1500 mL
And:                        
VTH                                                    250 mL ~ 1500 mL   
VTL                                                  0,20 mL ~ 300 mL
  1. Minute ventilation(MV)                    
--Max MV                                   
VTH                                                   18 L/min
VTL                                                     10 L/min
  1. PEEP
--Adjusting range                                0 kPa ~ 3kPa
-- Allowable error                        ±15 % or ±(0.3 kPa+10 % setting valve)whichever is the greater
  1. Pressure trigger(PTr)          
--Adjusting range                              - 2.0 kPa ~ 0kPa(Under PEEP)
--Allowable error                            ±(0.1 kPa+10 %setting valve)
  1. Flow trigger(FTr)                        
--Adjusting range                                               OFF, 0.5 L/min ~ 30 L/min
--Allowable error                                  
                                            >3L/min:   ±20%
3L/min:  ±0.6 L/min
  1. Pressure control(Pc)                         
--Adjusting range                                   0.5kPa ~ 6.0 kPa
-- Allowable error                         ± 15 % or ±(0.3 kPa+10 % setting valve)whichever is the greater
  1. Holding time(Inhalation platform)(TIP)  
--Adjusting range                                               0 ~ 6s
(0% ~ 50% of inhalation time)
--Allowable error                    ± 15 % or± 0.1 s,whichever is the greater
  1. Sigh
     --Adjusting range                           0, 1/100 ~ 5/100
                                              (Sigh tidal volume is 1.5 to 2 times higher than setting valve of tidal volume)
  1. Apnea ventilation
--Adjusting range                            OFF,5 s ~ 60 s
  1. Max set working pressure(pressure limit range)                          
--Adjusting range                                                 3kPa ~ 10kPa
--Allowable error                                                ± 15 % or ±(0.2 kPa+10 % setting valve)whichever is the greater
Max pressure                                                        12.5 kPa
explanation:
1) set pressure limit value to produce maximum working pressure( the upper limit of high airway pressure alarm).
2) do not use negative pressure at exhalation.
3) maximum limit pressure is pressure of the safety valve.

Driven gas

  1. Gas supply                                                   O2(For medical use)
  2. Input pressure range                                 280 kPa ~ 600 kPa

Fresh gascompensation

  1. Compensation range                                    0 ~  15 L/min
  2. gas composition                               O2, N2O and anaesthetic
 

Ventilation parameter Monitoring

  1. The following are monitoring parameters under the environment of body temperature and pressure-saturated. 
  2. Minimum exhalation volume maybe monitored is 50 mL/min.
  3. Fig. display:
ParameterRangeResolving powerAccuracy
Frequency(Freq)1 bpm ~ 100 bpm1 bpm ±15 %
Tidal volume(VT)0~ 2000 mL10 mL>100 mL:±20%
100 mL:±0.02 L
Minute ventilation(MV) 0~100L/min0.1 L/min>1L/min:±20%
1L/min:±0.5 L/min
(Ppeak)0~100 cmH2O1 cmH2O±(4%full scale+4%actual reading)
O2 concentration15 % ~ 100 %0.1 %±[ 2.5 %(v/v)+2.5 %(concentration level)]
Compliance1~1000mL/ cmH2O1mL/ cmH2O 
  1. Waveform display:
--Time-airway pressure(under all modes)
--Time-flow(under all modes)
--pressure volume loop(under all modes)
--Time-volume(under all modes)
--Time-CO2(under all modes)
Remarks:The machine cannot record all the adjusting and monitoring values.

 