China Hospital Infant Radiant Warmer Neonate Bilirubin Phototherapy Equipment Baby Infant Incubator, Find details about China Infant Incubator, Baby Incubator from Hospital Infant Radiant Warmer Neonate Bilirubin Phototherapy Equipment Baby Infant Incubator
Hood | ||||
Tubing access ports | 10 | |||
Access door size | 18cm x 13cm | |||
Mattress to hood height | 48.0mm | |||
Soft bed mattress size | 73.6cm x38.6cm x1.8cm | |||
Temperature control modes: Skin and air temperature control mode | ||||
Air mode control temperature range | 20ºC-37ºC | |||
Air mode control override temperature range | 37ºC-39ºC | |||
Air mode control accuracy | ≤±1.0ºC | |||
Skin mode control temperature range | 34ºC-37ºC | |||
Skin mode control override temperature range | 37ºC-38ºC | |||
Skin temperature accuracy | ±0.3ºC | |||
Dual-skin temperature monitoring | Yes | |||
Servo Humidity | ||||
Humidity control range | 30%-95% RH | |||
Humidity control operating time without refilling | 24 hours | |||
Humidity control reservoir capacity | 1500ml | |||
Humidity display accuracy | ±5%RH | |||
Humidity control accuracy | ±5% RH | |||
Servo Oxygen (Optional) | ||||
Oxygen control range | 21 % to 65% | |||
Oxygen display accuracy of full scale | ±2% | |||
Oxygen control accuracy (100% calibration) | ±3% | |||
Oxygen control accuracy (21% calibration) | ±5% | |||
Oxygen display resolution | 1% | |||
Scale (Optional) | ||||
Weight range | 300g to 8kg | |||
Weight display resolution | 1g | |||
Weight accuracy | ±10g | |||
Operating Environment | 20ºC to 30ºC | |||
Humidity | 10 to 95% RH | |||
Air Velocity | Up to 0.3m/sec |
Features:
1. 4" LCD display
2. Temperature monitoring
3. Temperature servo control
4. Humidity Monitoring
5. Humidity Servo Control
6. Air temperature and body temperature contorl two modes
7. Mattress tilt ±12 °
8. Ergonomic designed handle with perfect height
9. Trend for 2, 4, 8, 12, 24 hours and up to 7 days
10. Front and rear double-walls and air curtain
11. Drawers Large (28.0cm x 24.7cm x 36.0cm)
12. Drawers Small (28.0cm x 11.3cm x 36.0cm) *2
13. Auxiliary Power Output*3
14. Backup battery to keep air flow and alarm after power failure for more than 30 mins