GD-261D High Precision Petroleum Products Pensky-Martens Closed Cup Flash Point Tester
I. Introduction
1. GD-261D Automatic PMCC Flash Point Tester is designed and manufactured as per standards GB/T 261-2008 Determination of Flash Point - Pensky-Martens closed cup method and ASTM D93 Standard Test Methods for Flash Point by Pensky-Martens Closed Cup Tester.
2. GD-261D Automatic PMCC Flash Point Tester is used to make determination of the closed cup flash point of the petroleum products in the temperature range from 40° C to 360° C.
II. Main Technical Features
1. It adopts LCD screen to display. Full English man-machine dialog interface. It can preset the parameters of expected flash point, sample mark number, atmospheric pressure, test date, etc. It has menu to prompt and input function of guide type.
2. It adopts simulation tracking display the function curve of temperature rising and test time. It has prompt functions of English mis-operation, test date and test time.
3. It is equipped with standard RS-323, 485 computer port. The lower computer can save 100 groups of historical data. The test data can be saved for a long time, transmitted and modified if the instrument being connected with computer.
4. It can correct the influence of atmospheric pressure and calculate the correction automatically.
5. Differential detection. Automatically correct the systematic deviation.
6. Automatically open the lid, ignite, detect and print test data. The test arm lifting up and down automatically.
7. It adopts electric ignition to fire light the gas flame. It will light only need to press the button.
8. Humanized design, beautiful and safe. Easy to operate.
III. Main Technical Specs
1 | Power supply | AC(220±10%)V,50Hz |
2 | Temperature measurement | Range: room temperature ~ 300 °C |
Repeatability :≤ 3 °C |
Reproducibility :≤ 6 °C |
Accuracy : 0.1 °C |
3 | Heating rate | Procedure A : (5 ~ 6) °C /min |
Procedure B : (1 ~ 1.6) °C /min |
Automatic control and manually adjustable. |
4 | Stirring rate | (1) Procedure A: (90 ~ 120)RPM |
(2) Procedure B: (250±10)RPM |
Automatic control and manually adjustable. |
5 | Igniting mode | Electric ignition. Gas flame diameter:3.2mm ~ 4.8mm |
6 | Working condition | ambient temperature : (10 ~ 40) °C |
relative humidity :≤ 80% |
7 | Maximum power consumption | 500W |
8 | Dimension | 520mm×360mm×310mm |
IV. Same Series PMCC Flash Point Tester
Item | GD-261 | GD-261B | GD-261A | GD-261D | GD-5208 |
Testing Range | ambient~300°C | ambient~250°C | ambient~250°C | ambient~300°C | -30~50°C |
Repeatability | ≤2 °C | ≤2 °C | ≤2 °C | ≤2 °C | ≤2 OR 3°C |
| | | | | |
Automatic up/down | No | No | Yes | Yes | No |
Sample | Standard oil cup | Standard oil cup | Standard oil cup | Standard oil cup | 2ml |
With micropriter | No | Yes | Yes | Yes | Yes |
Communication port | No | No | No | Yes | No |
Reproductility | ≤4 °C | ≤4 °C | ≤4 °C | ≤4 °C | ≤2 OR 3 °C |
Screen Display | No | No | Yes | Yes | Yes |