China NIBP Blood Pressure Module for Patient Monitor (B100), Find details about China NIBP Module, Patient Monitor Module from NIBP Blood Pressure Module for Patient Monitor (B100)
Measuring range | Adult: 30 to 255 mmHg systolic blood pressure An average of 20 to 235 mmHg pressure 15 to 220 mmHg diastolic blood pressure Pulse rate of 30 to 240 times / min Newborn: 30 to 135 mmHg systolic blood pressure An average of 20 to 125 mmHg pressure 15 to 110 mmHg diastolic blood pressure Pulse Rate of 40 to 240 times / min |
Measurement time | Adults: up to more than 120 seconds; Newborn: up to more than 90 seconds; |
Restriction on software over-voltage | Adult: Less than 290mmHg; Newborns: no more than 145mmHg; |
Protection on hardware over-voltage | Adult: 300 ±10 mmHg; Newborn: 150 ±4 mmHg; |
Measuring principle | Oscillometry |
Deflation approach | ladder Deflated |
The initial inflation pressure value | Adult: 150 mmHg; Newborn: 70 mmHg; |
Sensor zero reset | Prior to each measurement |
Resolution | 1 mmHg |
Measurement accuracy | Maximum average deviation: no more than ± 5 mmHg; maximum standard deviation: less than 8 mmHg; |
Pressure accuracy instructions | ± 3 mmHg |
Leakage | less than 6 mmHg Per minute; |
Electrical performance | Voltage: 12 VDC or 5 VDC (can be selected by user), Quiescent current: 80 mA Measurement current: 200 mA inflated current: 600mA |
Environment | Operating Temperature: 0 to 55 Celsius degrees Storage Temperature: -20 to 65 Celsius degrees Humidity Range: 0 to 95%, non-condenser Operation Altitude: -500 m to 4600 meters Life Span: can operate 100,000 blood pressure measurements |
Size | 7.3 cm x 12.7 cm x 3.3 cm |
Weight: | 180 g |