China Veterinary oximeter handheld Pulse Oximeter for Vet clinic hospital optional capnography, Find details about China Veterinary Oximeter, Veterinary handheld Oximeter from Veterinary oximeter handheld Pulse Oximeter for Vet clinic hospital optional capnography
uPM60V Veterinary Handheld Pulse Oximeter
Technical Specification | |||||
Enclosure | |||||
Weight | 265g Weight(with batteries installed) | ||||
Dimensions | 80mmWx160mmHx30mmD (3.1in.Wx6.3in.Hx1.2in.D) | ||||
Alarms | Yes | ||||
Categories | Patient status and system status | ||||
Notification | Audible and visual | ||||
Setting | Full storage, low battery and sensor off | ||||
Range | Pulse rate (30-350bpm), SpO2 (0-100%) | ||||
Alarm volume level | 0-85dB | ||||
Range, Accuracy and Resolution | |||||
Range Type | Range Values | ||||
Measurement Ranges | |||||
SpO2 measuring range | 0-100% | ||||
Pulse rate range | 30-350bpm | ||||
Measurement Accuracy | |||||
SpO2 measuring accuracy | 70-100% (±2%), <70% undefined | ||||
Pulse rate measuring accuracy | ±2bpm/±2% take maximum | ||||
Resolution | |||||
SpO2 | 1% | ||||
Pulse rate | 1bpm | ||||
Electrical | |||||
AC power requirement | 100-240VA, 47-63Hz | ||||
Power comsumption | 20VA | ||||
Batteries | |||||
Types | Alkaline/Ni-Mh Batteries, IEC: LR6 | ||||
Operating time | ≥20hours | ||||
Charge time | ≤ 6hours | ||||
Environment | |||||
Operating tempearture | 0-45ºC | ||||
Storage temperature | -20-60ºC | ||||
Operating humidity | 30-95% | ||||
Operating altitude | -500-5000m | ||||
Trends | |||||
Types | Tabular | ||||
Memory | Save total 120 hours data events. Save data and time, alarm conditions, pulse rate and SpO2 measurements |