Ce Approved Transformer Oil Breakdown Voltage Tester (IIJ-II-60KV)

China Ce Approved Transformer Oil Breakdown Voltage Tester (IIJ-II-60KV), Find details about China Oil Testing, Insulating Oil Analysis from Ce Approved Transformer Oil Breakdown Voltage Tester (IIJ-II-60KV)

Model NO.
IIJ-II-60
Trademark
TOP
Origin
Chongqing, China
HS Code
9027809900
Model NO.
IIJ-II-60
Trademark
TOP
Origin
Chongqing, China
HS Code
9027809900

CE Approved Transformer Oil Breakdown Voltage Tester (IIJ-II-60KV)

 

Product Application:

IIJ-II series fully automatic oil tester is apply to test the insulating oil's dielectric strength, which is designed as per the requirement of IEC156 Testing Method of Insulation Oil Medium Strength, which via proper programming, has got functions of auto boosting, step down, stirring, display and print out. After breakdown of oil medium, HV break off will operate within 10 seconds to enable an auto isolation of control system partly; this effectively safeguards the personnel security and equipment safety.

Main Features: 
The automatic oil tester adopt industrial SCM to control which is applied to integrate circuit, new model I/0 connection , particular testing and anti-jamming technology which is largely improved the machine's capacity .

1. Simply operate: automatically (can be finished just press on key)

2. Menu controlled. Input data can clear

3. Can store 99 groups' oil testing data which is convenient to reading and print. The testing data can keep 100 years under the condition of power off.

4. System time can go on running under power off; (The system time just for reference can not be the standard of timing)

5. Can upgrade software through RS232

6.High capacity of anti-jamming and high precision of testing.

7. Small volume, light which operated in the open air very conveniently.
 

Technology and parameter data

1. Power supply: AC220V ±10%, 50 HZ 

2. Output voltage: 60 kV, 80 kV, 100 kV (100 kV should be specially designed) 

3. Capacity: 1.2 kVA (for 60kV), 1.6 kVA (for 80 kV), 2.0 kVA (for 100 kV) 

4. Pressure increase speed: about 2 kV/S 

5. Voltage inspection accuracy:   3% 

6. Dielectric strength sensitivity: <2KV 

7. Wave from distortion:   < 3% 

8. Breakdown response time:   <10ms 

9. Polar distance:   2. 5mm 

10. Working environment: temperature: 0 to 40 º C, humidity: <85% (Relative Humidity)

11. Storage environment: temperature: -20 to 60 º C, humidity: <75% (Relative Humidity)

12. Working Altitude: < 1500 meters. (If over 1500m, it can be customized)

13. Dimensions and weight: 38 x 37 x 38 cm. Net: 40 kg (Approximately)

Ce Approved Transformer Oil Breakdown Voltage Tester (IIJ-II-60KV)


 

Our oil tester/ analyzer:  

1. Transformer oil Breakdown voltage Tester (BDV test kit) series IIJ-II

2. Acidity tester model ACD-3000I

3. Karl Fischer Water content Tester Model TP-6A / model TP-2100

4. Fully automatic Interfacial Tension Analyzer model IT-800

5. Gas-Chromatography transformer oil gas analyzer model DGA2013-1

6. Multi-functional Vibration tester model TYD-2000

7. Consolidation Point Determinator model PT-2000

8. Kinematic Viscosity tester model VST-8 /model VST-2000

9. Flash point tester (open cup) model TPO-3000

10. Flash point tester (close cup) model TPC-3000

11. TAN-DELTA TESTER, 10kV Capacitance and Dissipation Factor Test Set model CDEF

12. Automatic Freezing Point & Pour Point Tester model TP-501Z

13. Determination of foaming characteristic of lubricating oil model SD-12579

14. Earth resistance tester model HYDW-II

15. Coil Resistance Tester (TPZC-40)

16. CT/PT tester model TPVA-404

17. Transformer turn ratio tester model HYG-II

18. Relaying Protection Tester model TPJB-III

19. Digital Grounding Resistance Tester model TP2571-II

20. DC High Voltage Generator model ZGF

21. Underground water/gold detector model TP-UWD

If you would like to know more about our oil purifiers and oil testers, please feel free to contact us any time. due to a sales as I, it is my biggest pleasure and responsibility to help my old and new clients to buy the high quality goods with best prices.

Power supplyAC220V±10%, 50Hz
Output voltage0-80KV (60KV.80KV)
Capacity1.2KVA (1.6KVA.2.0KVA)
Speed of pressure riseabout 2KV/S
Voltage test precision±3%
Breakdown sensitivity<2KV
Wave difference<3%
Time of breakdown<10ms
Working altitude<1500m (if over 1500m can be designed specially)
Dimension480mm*460mm*460mm
Working condition
Temperature  30Celsius degree--40Celsius degree
humidityThe biggest comparatively humidity85%
Reserve circumstance
Temperature-20Celsius degree-60Celsius degree
humidityThe biggest comparatively humidity75%

Our Company
Chongqing TOP Oil Purifier Manufacture Co., Ltd is a professional leading manufacturer of oil purifiers and oil testers. We takes the mission of existing on the basis of excellent quality as well as developing technical innovation,and introduces advanced technology.

TOP Oil Purifiers are widespread in the fields of electric power, petrifaction, mineral, shipbuilding, steel construction, bridge, automobile and railway, playing a great role in reducing energy consumption for clients as well as environmental pollution. We provides you with reliable oil purifiers such as transformer oil purifier,turbine oil purification machine, hydraulic oil purifier, lubricating oil filtration machine, fuel oil restoration, oil and water separator, oil regeneration equipment, waste oil recycling equipment etc.

TOP Testers works on the development, production, market, technical and after-sale service for petrochemical analyzers, transformer substation electrical testing equipment and circuit detecting equipment. Main products are water content testers, surface/interfacial tension testers, flash point testers, kinematic viscosity testers, oil acidity testers, insulation oil dielectric strength testers, transformer characteristics testers, relay protection testers, high voltage testing equipment etc.
Ce Approved Transformer Oil Breakdown Voltage Tester (IIJ-II-60KV)

Pre-sale Service
1)Will give response within 24 hours for all inquiries and  questions about our products.
2)Recommend the most suitable machine according  customer's demands.
3)Professional pre-sale model/configuration selection advise  and technical support -- offered by experienced technician.
4)Free on-site installation, debugging and technical training --  offered by strictly trained technology application engineers.
5)Offer OEM service.

After-sale Service
1)The warranty period is 24 months from commissioning acceptance date on. After the warranty period is over we provide lifelong maintenance and technical service.
2)We guarantee the maintenance time no more than 3 working days and response time within 1 hour.
3)We build instrument service profile for our clients to record   the product service and maintenance conditions.
4)After instruments start service, we will pay follow-ups to collect the service conditions.

Packing and Delivery
Ce Approved Transformer Oil Breakdown Voltage Tester (IIJ-II-60KV)

Safety and Timely Transportation
1)Standard export wooden case with high quality cushioning material ensure the item arrive customer in good condition.
2)Carefully check every packing detail before delivery to avoid any troubles during the transportation.
3)Appointed logistics company could be used according to customers demands.
4)Do our best to make sure customer can get their goods in time.

Contact Us