Automatic Chemistry AnalyzerHP-CHEM300H
HP-CHEM300H Auto Biochemistry Analyzer is our new designed laboratory popular chemistry analyzer.Perfect designing with table top style,make it much fashion in your lab.And it can be produced by OEM style.
Enhanced Liquid System1.Precise sample and reagent aspiration |
2.Built in bubble elimination prior to washing |
3.High pressure interior probe wash |
4.Carry-over<0.05% |
Economy Usage1.Light-spot Flatting Technology facilitates lower reaction volume |
2.24 hour non-stop refrigeration at 4-10ºC |
3.Minimum sample volume:2ul |
4.Reagent volume:20-25ul |
5.Minimum cuvette volume:150ul |
Easy to perform maintenance1.Front loaded reagent compartment for easier access |
2.Easy access for routine maintenance and troubleshooting |
3.Easy to connect pipeline and cable |
User-friendly Interface1.Real-time analytical and carousel status monitoring |
Real-time QC Status Monitoring1.Levy-Jennings chart and Twin-plot chart |
2.QC Out-of-Range real-time alarm |
Traceable Test Results1.Historic date recall from reagent/calibrator/control archive |
2.Intuitive software,easy historical results |
Reflex function1.User-defined customizable reflexive assays |
2.Multiple reflexive criteria may apply to a single assay |
3.A single reflexive criteria may apply up to 20 assays |
Technical SpecificationsSystem Function |
Fully automated,discrete,random access,STAT,urine and homogeneous immunoassays,STAT sample priority |
Throughput | Constant 300 tests/hour,Massuring principles:Absorbance Photometry,Turbidometry |
Methodology | End-point,Fixed-time,Kinetic,Single/uual/Triple/Quadruple reagent chemistries,Monochromatic/Bichromatic |
Programming | User defined profiles and calculation chemistries |
Sample Handling |
Sample tray | 120 positions for sample primary or secondary tubes and sample cups |
Sample volume | 2~100ul,step by 0.1ul |
Sample probe | Liquid level detection,clot detection and collision protection |
Probe cleaning | Interior and exterior automatic probe washing carry-over<0.05% |
Reagent Handling |
Reagent tray | 80 positions in refrigerated compartment (4~10ºC) |
Reagent volume | R1:150~300ul,R2:20~150ul |
Reagent probe | Liquid level detection, collision protection and inventory check |
Probe cleaning | Interior and exterior automatic probe washing |
Reaction System |
Reagent rotor | Rotating tray,81 cuvettes with automatic washing |
Cuvette | Optical length 5mm |
Reaction volume | 150~450ul |
Operating temperature | 37ºC |
Temperature fluctuation | ±0.1ºC |
Mixing system | 2 independent mixers |
Optical System |
Light Source | Halogen-tungsten lamp |
Photometer | Reversed optics photometry |
Wavelength | 340nm 405nm 450nm 510nm 546nm 578nm 630nm 670nm |
Absorbance range | 0~4.0Abs(10mm conversion) |
Resolution | 0.0001Abs |
Control and Calibration |
Calibration mode | Linear(one-point,two-point and multi-point),Logit-Log 5p,Spline,Factor |
Operation Unit |
Operation system | Windows XP SP2 or SP3/ Windows 7 Home Basic |
Interface | RS-232 |
Working Conditions |
Power Supply | 200~240V 50/60HZ 1000W or 100~120V 60HZ 1000W |
Temperature | 15~30ºC |
Humidity | 35~85% |
Water consumption | 18L/hour,De-ionized water |
Dimension | 912mm*721mm*570mm(W*D*H) |
After Sale and Complaint
1.After-sale Service Policy
1)All our products are one year quality guarantee time generally.Longer guarantee can be refer to detailed offer.
2)We will supply spare parts supports for most of our products for 5 years at least,and we will inform updated information for products supplied or stop production to our agent and showed them in our official website.
3) If any quality problems such as design, manufacture, function or procedure occur within the guarantee period,we will in charge to maintain or supply spare parts for the equipment in reasonable ways or replacement the equipment if it certified to be needed.
4)The cost for spare parts is free and labor is free too within guarantee period.And it will be charged by reasonable price out of guarantee period.The delivery fees,traffic frees and subsistence fees are charged accordingly or supplied by your side within and out of guarantee time.
5)The primary positive response should be supplied through phone or email or other reasonable ways within 24 hours after getting the notice from agent or end users.And the details settlement program should be supplied by paper or email in days no more than 6 working days after getting the After-sale support from.
6)We will provide a lifetime paid maintenance service and technical support for all our products out of guarantee period.And we will inform customers the update information for our products or suggest to clients to order the new model.
7)End users or agent can make phone or fax or email for getting after-sale service for emergency and then fill the After-sale Support Form and send it to us by email or fax in time.
8)HEPO INTERNATIONAL has a self-governed team for deal with after sales or complain business.
2,Complain Policy
1)End users or agent can make phone or fax or email for complain for emergency and then fill the Complaint Form and send it to us by email or fax in time.The positive response time is same as upon clause in after-sale service.
2)Customs' questions is first.If the goods received was breakdown or the package was damaged,we can arrange the new goods at once getting the inquiry from agent or purchaser along with pictures or other needed certified documents in our side.And we will bear half freight of sea shipment of this new order without any complain.The duty for the breakdown or damage of the head one can be deal after this new delivery.
3)Once we delay delivery date,we will bear half freight for sea shipment for this order without any complain.