China Full-Scale Indoor Fire Test Device That Complies with ISO 13784-1 and ISO 9705 Standards, Find details about China Full Scale Room Test Chamber, Full Scale Room Fire Tester from Full-Scale Indoor Fire Test Device That Complies with ISO 13784-1 and ISO 9705 Standards
1.The test room is built by the density of 700kg / m³-A1 class fireproof material.
2.The internal dimensions of the room 3.6M X 2.4M X 2.4M
3.The gate size: 0.8M (W) X2.0M (H)
4.Sandbox burner with propane gas and equipped with emergency shut-off devices.
5.Full Scale Room Fire Test Chamber adopts automatic ignition system to automatically record experimental data;
6.Independently movabletimetto display units.
7.10kV spark ignition, a cease-fire with safety devices. Fire extinguishing drainage system design.
8.The full scale room fire test machine is equipped withstaircasefor easy maintenance exhaust pipes.
1, Gas sampling system includes an annular sampler probe, suction pump, particulate filter, a cold trap, exhaust valve, water filters and CO2 filter.
2, Full Scale Room Fire Tester is supplied with three filters; Minimum filtration accuracy is 0.5um.
3, Freezing cold trap; air-conditioning compressor refrigeration systems; controlling the temperature in 0 ~ 5ºC.
4, Suction pump,Flowrate is 33L/min,vacuumis 700mmHg, and pressure is 2.5 bar.
5, Full Scale Room Fire Test Equipment has drying cylinder, which filter the water and impurities.
6, The flow controller to control the inspiratoryair flow.
7, The sampler has 12 holes and gas flow in the opposite direction, these holds built-in dust filter and 685mm from hoods.
8, Paramagnetic oxygen analyzer, the range is 0-25%; linear response; imported Britain Servomex oxygen sensor.
1, 0.5W HE-NE laser source.
2, Thorlabs optically detector.
3,Motor driven 5 class automatically calibration device with of 0%, 25%, 50%, 75%, 100% transmittance filters.
4,One touch calibration system in software with digital and curve display.
1, 16bit data acquisition module.
2, Digital communication between computer and DAQ module.
3, Labview based application software on DAQ and data analysis.
1, Labview based application system,easilyand precise to operate and control.
2, Test status, calibration status display onscreen.
3, Device status are displayed on screen and indicators onpanel.
4, Transducers data, process parameters, test result displayed onscreenvia digital and curve.
5, Test and calibration procedure auto executed byPLCprogram.
6, Full Scale Room Fire Testing Equipment can be controlled via both application screen and control panel.
7, PC and PLC communicate with digital protocol which is safe and robust.
1,Paramagnetic oxygen analyzer, the concentration range is 0~25%, an oxygen analyzer islinearresponse, Britain Servomex oxygen sensor; 10% to 90% response time of less than 12s; output noise does not exceed 50ppm;.Reproduce the measured values lower than 100ppm,
2, Infrared CO2 analyzer range: 0 to 10%;accuracyof ± 0.01%;
3, The compressor works: cold trap: 0 to 5 degrees;
4, Full Scale Room Fire Test Machine uses imported diaphragm pump, flow rate: 33 l / min
5, The standard detector color accuracy ± 5%, output linearity (transmission) <1%, absolute permeability <1%;
Test Data
1, Heat Release Rate (HRR),
2, Total Heat Release (THR),
3, Efficiency Heat Release (EHR),
4, Smoke Production Rate (SPR),
5, Time to Ignite (TTI),
6, Time to flameout,
7, Oxygen Consumption Rate,
8, CO2 Production Rate,
9, CO Production Rate(optional)
Size | 11,725(W)x4,000(D)x4,750(H)mm |
Controller Size | 600(W)×750(D)×1800(H)mm |
Power supply | AC 220V, 50/60Hz, 15A (control), AC 380V, 60Hz, 50A(drive) |
Weight | About 7,500kg(Cabinet), 130kg(Controller) |
Exhaust | Minimum 3.5m2/s(Measuring), Minimum 300m2/min(Exhaust) |
Other requirements | CO(0.8%)/CO2(8%), Compressed gas, Vacuum cleaners, Computers, Propane gas, Heptane gas, Nitrogen |