Biobase China 100 Liter Horizontal Pulse Vacuum Steam Sterilizer Autoclave, Find details about China Vacuum Steam Sterilizer, Horizontal Autoclave from Biobase China 100 Liter Horizontal Pulse Vacuum Steam Sterilizer Autoclave
Application:
It can be widely used in department of stomatology and ophthalmology, operating room and other medical institutions. It is suitable for all wrapped or unwrapped solid instrument, A-class cavity instrument (dental hand-pieces and endoscopes), implantable instrument, dressing fabric and rubber tubes, etc.
Features:
1. Built-in efficient steam generator, microcomputer temperature control.
2. Equipped with printer to print date, time and process parameters.
3. Prefect four-level authority management system to avoid mis-operation.
4. LCD displays temperature, pressure, time, running status, failure and warning code.
5. Sterilization programs for fabric, instrument, rubber, liquid, cavity, B&D test, PCD test and leakage test.
6. Intelligent drying system to guarantee the residual humidity of instrument and dressing no more than 0.2%
and 1% respectively.
7. Overheat auto protection device, dual door safety lock device, over pressure relief valve, short circuit protection,
and overpressure protection.
Technical Parameters:
Model | BKQ-Z100(H) | BKQ-Z150(H) | BKQ-Z200(H) | BKQ-Z300(H) | |||
Capacity | 100L | 135L | 185L | 300L | |||
Chamber Size | Φ400*850mm | φ500*710mm | φ500*950mm | φ632*1000mm | |||
Chamber material | 304 medical stainless steel | ||||||
Max working pressure | 0.23Mpa | ||||||
Vacuum limit | -0.08MPa | ||||||
Working temp | 105~136ºC | ||||||
Temp accuracy | ±0.1ºC | ||||||
Power consumption | 11kw | 17kw | |||||
Power supply | 380V, 50HZ | ||||||
External size(mm) | 1250*730*1660 | 1110*750*1742 | 1350*750*1742 | 1400*890*1780 | |||
Packaging size(mm) | 1450*1000*1950 | 1370*890*2060 | 1510*890*2060 | 1700*1050*2090 | |||
Gross weight(kg) | 310 | 470 | 540 | 830 |