China Contec Cms5000vet Veterinary Animal Hospitals Vital Sign Patient Monitor Vet Instrument, Find details about China Vet Instrument, Patient Monitor Vet from Contec Cms5000vet Veterinary Animal Hospitals Vital Sign Patient Monitor Vet Instrument
NIBP monitoring | Measurement method: Oscillometry Measurement mode: manual/auto. Auto measurement interval: adjustable from 1 to 255 minutes. Resolution: 1mmHg. Accuracy: Maximum Mean deviation±5mmHg, Maximum Standard deviation≤8mmHg. Self-check function. Overpressure protection: double protection for software and hardware. Measurement range: 10mmHg~270mmHg. |
SpO2 monitoring | Continuous real-time measuring. Adopt digital SpO2 technology, which has strong anti-interference capability. Real-time display SpO2, pulse rate, SpO2 Plethysmogram, bar graph, SpO2 and Pulse rate (PR) trend graph. |
Trend graph with the function of adaptive scaling | |
1.SpO2 Measuring | 1.Range: 0%~100% (Resolution: 1%) 2.Accuracy: ±2% in stage of 70%~100%, and unspecified in stage of less than 70% |
Pulse Rate Measuring | 1.Range: 30bpm~250bpm(Resolution: 1bpm) 2.Accuracy: ±2bpm or ±2%, whichever is greater |
Measurement Performance in Weak Filling Condition | SpO2 and pulse rate can be shown correctly when pulse-filling ratio is 0.4%, SpO2 error is ±4%; the Pulse rate error is ±2bpm or ±2%, whichever is greater |
Resistance to surrounding light | The deviation between the value measured in the condition of man-made light or indoor natural light and that of darkroom is less than ±1% |
Operational environment | |
Temperature | 5 ˚C~40 ˚C |
Relative humidity | 30%~75% |
Atmospheric pressure | 700hPa~1060hPa |
Specific EMC, climate, mechanical environment: Do not use a mobile phone in the vicinity of the veterinary monitor for the strong radiation field produced by mobile phone would interfere with the normal function of the veterinary monitor use. |
-TELEMEDICINE products