TBT-265E Asphalt Kinematic Viscosity Tester

China TBT-265E Asphalt Kinematic Viscosity Tester, Find details about China Kinematic Viscosity Testing, Petroleum Viscosity Tester from TBT-265E Asphalt Kinematic Viscosity Tester

Model NO.
TBT-265E
Viscometer Reproducibility
±0.2%
Viscometer Rotor Number
6
Viscometer Rotor Speed
60r/min
Viscometer Rotor Specifications
0 Size
Ambient Temperature
180.0oc
Temperature Controlling Accuracy
± 0.1oc
Bath Cubage
Not Less Than 23 L
Stirring Motor
6 W, 1200 Rpm
Trademark
TBT
Transport Package
Standard Export Packing
Specification
ISO9000
Origin
Nanjing, China
HS Code
9031809090
Model NO.
TBT-265E
Viscometer Reproducibility
±0.2%
Viscometer Rotor Number
6
Viscometer Rotor Speed
60r/min
Viscometer Rotor Specifications
0 Size
Ambient Temperature
180.0oc
Temperature Controlling Accuracy
± 0.1oc
Bath Cubage
Not Less Than 23 L
Stirring Motor
6 W, 1200 Rpm
Trademark
TBT
Transport Package
Standard Export Packing
Specification
ISO9000
Origin
Nanjing, China
HS Code
9031809090

 TBT-265E Kinematic Viscometer
TBT-265E Asphalt Kinematic Viscosity Tester


The instrument is designed and made as per the national standard of People's Republic of China GB/T1841 Standard Test Method for Viscosity of Polyolefin Resin Solution. 
It is used to determine the viscosity of decahydronaphthalene solution of polythene and polypropylene at the stipulated temperature. The instrument can also be used to determine the viscosity of asphalt at the stipulated temperature as per the industrial standard of People's Republic of China T0619 Asphalt Kinematic Viscosity Test (Capillary Viscometer Method).
 
  1. Main technical features
    1.High precision digital temperature controller. Electric stirring motor.
    2.High temperature resistant glass bath. Double layer structure.
    3.Desktop structure. All-in-one design.
    4.The maximum heating temperature is 180ºC.
 
  1. Main technical specifications
Power supplyAC 220 V±10%, 50 Hz
Heating power1700 W
Temperature rangeAmbient to180ºC
Temperature control accuracy±0.1ºC
Mercury-in-glass thermometerScale division 0.1ºC. 
Range 100ºC~150ºC and range 150ºC~200ºC(for maximum 180.0ºC)
Bath capacityabout 23 L
Sample quantityYou can make determination using 3 capillary viscometers at a time
Stirring motor(1) Power: 6W  (2) Speed: 1200 RPM
Ambient temperature-10ºC~+35ºC
Relative humidity<85%
Temperature sensorRTD, Pt100 
Maximum power consumption1800 W
Capillary viscometerOne group of Cannon-Fenske Opaque capillary viscometers; 7 pieces in total. They are N0.200, 300, 350, 450, 500 and 600      (The inner diameters of R tube of them are 1.02, 1.26, 1.48, 1.88, 2.20, 3.10 and 4.00 mm respectively).
Overall dimension530mm×400mm×670mm(Bath is included)

III. Standard accessories
No.ItemUnitQtyRemarks
1Countercurrent capillary viscometer
(a group)
Piece7 
2Mercury thermometer (100~150ºC, scale division is 0.1ºC)Piece1Standard configuration
Mercury thermometer (150 ~200ºC, scale division is 0.1ºC)Piece1According to the users requirements to purchase
3Weight ballPiece1 
4SensorPiece1 
5Sensor sheathPiece1 
6Bakelite coverPiece4 
7Stopping ring for thermometer (1#)Piece2 
8Capillary viscometer holderPiece4 
9Silica gel stopper (3#)Piece2 
1015A fuse  (Ф6×30)Piece4 
11Round glassBlock1 
 
IV.TBT-265E Kinematic Viscometer's Reagent and Auxiliary Equipments 
No.NameSpecificationQuantityMemo
1Glycerol or silicone oilOpen flash point >215ºC20LWhen test temperature above 180ºC,using silicone oil
2Petroleum solution1L1L/bucketTrichloro ethylene.gasoline,kerosene,choose one only
3Distilled waterThree grade1 litre 
4Washing ear ball 1 piece 

Notes:
1.The reagent and equipments for installing the instrument are supplied by end-users themselves.
2.The usage,management and saving of poisonous and harmful reagent,pls operate according to the relative principle.
3.PLS read the operation manual and industry standard before using the instrument.