Name: Multiparameter veterinary monitor
Model No: BW3M-VFeatures1. Attractive streamline outline design, portable, compact and lightweight
2. 12.1" color TFT LCD display3. Operation menu with multi-language interface selection: English, Spanish, Portuguese, Chinese, French, German
4. Big Font Display
5. ST segment Analysis and Apnea Alarm function
6. Drug Dose Calculation and OXCRG Software
7. 72-hour tabular and graphic trends and data storage
8. High energy Li-ion battery for short-term transport and temporary power-down
9. Networking Capability. WAN communication function to network with central monitoring system and make long-distance monitoring, diagnosis, maintenance and software upgrade possible
10. Linux Operation System
11. External display(VGA interface) and external printer connectible to record tabular and graphic trend
12. Dynamic Monitoring of blood pressure
13. Intelligent audio and visual alarm
14. Application: Horses, Cats, Dogs or Similar Sized Animals
15. Pacemaker detection and Arrhythmia Analysis( Optional)
Technical SpecsECGInput: 5 lead ECG cable
Lead section: I, II, III
Gain(mm/Mv): 1/4, 1/2, 1, 2, 4
Frequency response:
Diagnosis 0.05 - 100Hz
Monitor mode 0.5 - 40 Hz
Surgery 1.0 - 25 Hz
Sweep speed (mm/sec): 12.5, 25, 50
Heart rate range: 0-350 BPM
Heart Rate Accuracy: ± 1%
Protection: Against electrosurgical interference and defibrillation
NIBPMethod: Automatic Oscillometric
Operation Modes: Manual/Automatic/Continuous
Measurement Unit: MmHg/kPa selectable
Measurement range: Systolic, Diastolic and Mean
Measurement range: 10-285mmHg
Scale: MmHg/Kpa
Accuracy: ± 3mmHg or 2%(choice max)
Alarm: Systolice, Diastolic and Mean
Over-pressure Protection: Yes
Resolution: 1mmHg
SpO2Measurement range: 0 - 100 %
Pulse Rate range: 25~400BPM
Accuracy: 70~100% ± 2%(Horse & Dog)
70~100% ± 3%(Cat)
0~69% Unspecified
Low Perfusion: < 0.3%
Low Perfusion Accuracy: 70~100% ± 2%(Horse & Dog)
70%~100% ± 3%(Cat)
0%~69% Unspecified
TEMPERATUREAlarm Range: 27~42celsius
Measurement range: 25~45celsius or 77~113fahrenheit
Accuracy: ± 0.2celsius or± 0.3fahrenheit
Scale: Celsius/Fahrenheit
RESP.Method: Impedance
Measurement range: 0 - 120 BPM
Accuracy: ± 1BMP or 5%(choice max)
Resolution: 1BMP
POWER REQUIREMENTInput: 98~260 V AC, 50/60Hz
Consumption: ≤ 80VA
Safety standards: GB9706.1 IEC 60601-1 UL 544